Agilent Technologies N5242A Service Manual page 308

2-port and 4-port pna microwave network analyzers (10 mhz–26.5 ghz)
Hide thumbs Also See for N5242A:
Table of Contents

Advertisement

General Purpose Maintenance Procedures
Error Terms
Figure 8-3
Flowgraph of Three-Port Error Terms
where:
E = error term
DIR = Directivity
MAT = Forward Source Match and Reverse Load Match
TRK = Forward Reflection Tracking and Reverse Transmission Tracking
For the case of a full 3-port calibration, port 1 has
three Match error terms:
S11 source match
S12 load match
S13 load match
and three Tracking error terms:
S11 reflection tracking
S12 transmission tracking
S13 transmission tracking
There are six isolation terms not shown.
8-6
PNA Series Microwave Network Analyzers
Service Guide N5242-90001
N5242A

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents