Invacare® SB®755

4 Maintenance

4.1 Cleaning and disinfection

IMPORTANT!
Wrong fluids or methods can harm or damage you
product.
– Follow instructions carefully for either Non-washable
or Washable components.
– Never use corrosive fluids (alkalines, acid, cellulose
thinner, acetone etc)
– Never use a solvent that changes the structure of the
plastic or dissolves the attached labels.
– Always make sure that the bed is carefully dried before
taking it into use again.

4.1.1 Non-washable components

IMPORTANT!
Non-washable components cannot withstand washing
with Jet based equipment.
– The washability of the components must be determined
by authorized personnel.
– If the components cannot withstand Jet base
equipment or if it is unclear they have to be cleaned as
non-washable components.
12
Electrical components
Method:
Wipe off with a wet cloth or soft brush.
Max temp:
40 °C
Solvent/
Water
chemicals:
Metal components
Method:
Wipe off with a wet cloth or soft brush.
Max temp:
40 °C
Solvent/
Household detergent or soap and water, 6-8 pH
chemicals:
Wood (including textile straps on side rails, if existing)
Method:
Wipe off with a wet cloth or soft brush.
Max temp:
40 °C
Solvent/
Household detergent or soap and water, 6-8 pH
chemicals:
Textiles (including upholstery and mattresses)
Method:
See attached label on each product.
1567446-E