Figure 2-1. Jumper J7 Configuration; Figure 2-2. Jumper J6 Configuration - Omega Engineering DAQ-16 User Manual

Data acquisition adapter for 16-bit isa compatible machines
Table of Contents

Advertisement

Figures 2-1 and 2-2 show the configuration options for jumpers J7 and J6.
WARNING: These settings are only suggestions, it is the user's responsibility to guarantee
that the maximum input voltage multiplied by the gain setting selected by jumper J7 does
DAQ-16 Users Manual
J7
3
4
3
1
2
1
Gain = 1

Figure 2-1. Jumper J7 Configuration

J6
4
5
6
1
2
3
10 volt range

Figure 2-2. Jumper J6 Configuration

not exceed the A/D voltage range set by jumper J6.
J7
3
4
1
2
Gain = 10
J6
4
4
5
6
1
2
3
1
5 volt range
2.5 volt range
J7
4
2
Gain = 100
J6
5
6
2
3
11