Omega Engineering DVG64 Instruction Sheet

Electronic vacuum gauge m3805/0502

Advertisement

Quick Links

DVG64
Electronic Vacuum Gauge
INSTRUCTION
SHEET
Display
Sensor Type
Connector Type
Vacuum Range
Scale
Operating Temp. Range 35°F to 125°F ( 2°C to 52°C)
Overpressure
Accuracy
Power Source
Continuous Usage
Auto Shut off
Weight
Dimensions
Vacuum (microns)
15000 - 8000
8000 - 5000
5000 - 2000
2000 - 1000
1000 - 500
500 - 200
200 - 0
FEATURES
• Reads vacuum in 7 international units
• Uses Ω the battery power than other models
• Cleaning port provides quick cleaning and drying of sensor
• Uses standard 9V Alkaline battery
• Convenient built-in hanger
OPERATING CONTROLS
• Turning the vacuum gauge ON: Press and hold the ON button
for about 3 seconds, until "POWER UP" appears on the display.
• Turning the vacuum gauge OFF: Press the OFF button
To prolong battery life, when vacuum reading is above 12,000
Microns for about 10 minutes, the unit will automatically turn OFF.
• Changing the scale: Press the Scale button to change the display
to the next scale. The scale order is: Microns, PSI, Inches of
mercury (InHg), milliBars, Pascals, Torr and milliTorr. DVG64 will
keep the scale settings even if the power is turned OFF.
UNDERSTANDING THE DISPLAY
• When the vacuum reading is above 12,000 Microns (1,600
Pascals), the first line of the display shows "Atm." The second
line displays a bar graph to indicate the direction in which the
vacuum is moving. When the bar graph is moving from left to
right, the pressure is increasing. When the bar graph is moving
from right to left, the pressure is decreasing. The speed of the
bar graph indicates how fast the pressure is increasing or
decreasing. The bar graph indicator may be inaccurate for a few
seconds after the evacuation of the system has begun.
M3805/0502
DVG64 SPECIFICATIONS
Updated every Ω second. 1 micron resolution below 200 micron of vacuum
Thermistor
Standard 1/4 inch male flare fitting
0 - 12,000 Microns (0 - 1,600 Pascals) with vacuum increasing/decreasing indicator when above
12,000 Microns
Microns, PSI, InHg, milliBars, Pascals, Torr, milliTorr
300 PSI max (20 Bar)
+/-10% (0 to 1000 microns)
9 Volt Alkaline battery (not included)
Over 35 hours
After 10 minutes when vacuum reading is above 12,000 Microns (12 Torr)
6.7 ounces
5-1/2"H X 3"W X 1-1/4"D
DVG64 STEP TABLE
Micron (mTorr) Step
1000
500
250
50
10
5
1
Pascal Step
Bar Step
100
1
50
0.5
25
0.25
5
0.05
1
0.01
0.5
0.0005
0.1
0.0001
• The bar graph disappears if the vacuum does not change for
about 10 seconds. When the vacuum reading is below 12,000
Microns (1,600 Pascals), the vacuum in the selected units is
displayed.
CONNECTING TO THE VACUUM SYSTEM
The VG64 should only be connected to the vacuum system at
the vacuum port. The "Auxiliary Port" is primarily for cleaning
and should normally be closed with the supplied cap.
CLEANING THE VACUUM SENSOR
It is recommended that the sensor be cleaned periodically to
maintain unit accuracy . Oil and other contaminants reduce the
accuracy of the unit. Follow the instructions below for cleaning.
• Close the Vacuum port with the supplied cap.
Open the Auxiliary port.
• Use an eyedropper to pour about 2 teaspoons of ordinary
rubbing alcohol into the Auxiliary port.
• Close the Auxiliary port with the supplied cap. Both the
Vacuum and Auxiliary ports should now be closed.
• Shake the DVG64 unit for about 10 seconds. A slight
movement of the vacuum sensor in the case is normal and
does not affect the internal connection in any way.
• Open both the Vacuum and the Auxiliary ports.
Empty the alcohol, and air dry the sensor.
• Close both the Vacuum and the Auxiliary ports with the
supplied caps when the DVG64 is not used. This prevents
contamination of the sensor.
InHg Step
PSI Step
0.05
0.02
0.02
0.01
0.01
0.005
0.002
0.001
0.0005
0.0002
0.0002
0.0001
0.00005
0.00002

Advertisement

loading

Summary of Contents for Omega Engineering DVG64

  • Page 1 • Close the Auxiliary port with the supplied cap. Both the Vacuum and Auxiliary ports should now be closed. • Shake the DVG64 unit for about 10 seconds. A slight movement of the vacuum sensor in the case is normal and does not affect the internal connection in any way.
  • Page 2 WARNING: These products are not designed for use in, and should not be used for, patient-connected applications. OMEGA ENGINEERING, INC. warrants this unit to be free of defects in materials and workmanship for a period of 13 months from date of purchase. OMEGA’s WARRANTY adds an additional one (1) month grace period to the normal one (1) year product warranty to cover handling and shipping time.