GE DewPro MMY31 Installation And Operation Manual page 13

General eastern trace moisture transmitter
Hide thumbs Also See for DewPro MMY31:
Table of Contents

Advertisement

September 2005
Mounting the MMY31
(cont.)
2-2
Mount the MMY31 using a compression fitting connection. Be
certain that the tip of the probe does not touch the inside wall of the
pipe.
IMPORTANT:
Ensure that your probe has a stainless steel ferrule,
which is required for use in pressurized systems.
Adjust the ferrule for a probe insertion length of 1" (minimum) and
tighten the compression fitting as follows (see Figure 2-2 below):
1. Hand-tighten the nut.
2. Then, using a wrench, tighten the nut 1½ turns. Now the fitting is
tight and can withstand pressure to 1750 psig (120 barg).
Figure 2-2: Mounting the MMY31
Minimum 1 in. (25 mm)
Installation