Tec 6 Plus Vaporizer
Alarm and display test
À WARNING
À CAUTION
Step 1
Connect the power cord from
the vaporizer to an approved
hospital grade outlet socket.
À WARNING
Step 2
Verify that the alarm and
display test for a period of
approximately four seconds as
follows:
5-2
Each light and alt the LCD
agent level indicator bars
on the front display panel
flash.
The auditory alarm is
activated.
Only use vaporizers that are in a serviceable condition.
Before connecting the vaporizer to an AC power supply,
be sure it is within the ambient operating temperature
range (see Performance in Chapter 9).
Do not use the vaporizer if during the alarm and display
test any one of the five lights and all the LCD agent level
indicator bars do not flash or the auditory alarm does not
operate for a period of approximately four seconds.
D
DESFLURANE
バレ
DOL
OPERATIONAL
ALARM
BATTERY LOW
ω
ο.
Ιου
Figure 5-1 ¢ Verifying the alarm
Flashes
1/23/01
1107-0641-000