Advertisement

Deactivating alarms

You can set the alarm limits temporarily to OFF for the following parameters: HR, SpO
respiration rate. This way, you can silence a parameter without having to adjust the alarm
limits.
1.
Press the
2.
Select
3.
Turn the ComWheel to highlight the measurement.
4.
and push the ComWheel to change the selection
The symbol
When you select the alarm limit ON again, the previous alarm limits will be active.

Automatic recording on alarms

An automatic local recording is possible when Asystole, Tachy/Brady, Art High/Low, V Fib or V
Tachy alarms reach the red alarm level. When the red alarm level is reached, the recorder
prints the ECG + Art waveforms.
1.
Press the
2.
Select Record Waveforms.
3.
Select Start on Alarms and Yes.
Print/
Record
Alarms Setup
key.
Adjust Limits
.
appears in the digit field.
Print/Record
key.
Off
.
Alarms
and
2
4-11