CONTROL BUTTONS
The AED has two buttons.
SHOCK BUTTON
The SHOCK button is located at the far right of the control panel.
Delivers a defibrillation shock. The word SHOCK and the shock button indicator LED will illuminate
RED when the AED is ready to deliver a defibrillation shock to the patient. Note modification to
behavior below when in manual mode.
MANUAL OVERRIDE BUTTON
The MANUAL OVERRIDE button is located at the far left of the control panel and converts the
device from automated mode to manual. This feature should only be used by ALS personnel.
MANUAL OVERRIDE
• Lift the cover to access the button.
• Converts to manual standby mode when pushed once, a voice prompt "Press Manual Button Again
to Confirm", will be heard. Converts to manual mode when MANUAL button is pressed again.
• If the rescuer does not confirm within 30 seconds of the capacitors charging, the AED will revert
back to AED Mode.
• If the Medical Director has disabled this feature in Servicelink, an icon indicating No MANUAL
MODE will appear in the bottom left of the display.
REMAIN IN MANUAL MODE
This feature is intended for advanced medical personnel with the ability to differentiate between
shockable and non-shockable cardiac rhythms. With Remain In Manual Mode enabled and the user
enters manual mode, the AED will remain in the manual mode. Further AED rhythm analysis, and
CPR prompting are disabled when "Remain in Manual Mode" is enabled.

SETTING THE AED INTERNAL CLOCK

The internal clock is preset at Central Standard Time and should be reset to the correct date and local time. The AED will
automatically adjust itself for daylight savings time. This feature can be turned off using the ServiceLink software. To set
the clock, you will need a PC with Windows 95 or later operating system, RescueLink software installed, an IR port on the
PC, and an IR cable as specified below.
To set the clock settings:
Open the lid and remove pads from the pads socket.
Ensure that the PC is set at the correct local time and date.
Point IR port on the AED to IR eye on the PC and select G3 Pro.
Run the RescueLink software on the PC.
Verify that the voice prompt states "Communications Mode".
Click Communications on the main menu. Select AED Date and Time.
Click on the Get button to review the current time in the AED.
If the time and date is incorrect, click Set to set new time and date. The AED date and time will
automatically be updated to the PC's time and date.
Reinstall pads per instructions on page 19.
Close the lid.
Note: The IR port on the AED is designed to work with IR cable ACT-IR220LN115 from ACTiSys Corp. on
Windows based PCs only. Please contact customer service to order, P/N 162-0108-001. Other IR products
may interfere with the transmission and are not for use with the AED.