GE LOGIC C2 Series Technical Publication page 112

Ultrasound system
Table of Contents

Advertisement

Care and Maintenance
Isolated lead (sink) leakage test record (continued)
NOTE:
Table 4-14:
Typical data format for recording isolated patient lead (sink) leakage
Unit under test_________________________________
System Power
on
off
NOTE:
Table 4-15:
Typical data format for recording isolated patient lead (sink) leakage
Unit under test____________________________________
System Power
off
on
off
on
4-34
Keep a record of the results with other hard copies of
maintenance data using Table 4-14.
Not all test procedures are applicable to all areas of the world.
Reversed Polarity testing content satisfies regions following IEC
62353:2007 and IEC 60601-1:2005.
Test Conditions
Grounding/PE
closed
closed
Values in italics font are given as examples only.
Test Conditions
Grounding/PE
closed
closed
closed (reversed polarity)
closed (reversed polarity)
Ultrasound System
Date of test:___________
Patient Lead
Date of test:_____________
Limit µA
50
50
50
50
Common Service Information
Direction 5444964-100 English Rev. 5
RA+LA+LL
Patient Lead
RA+LA+LL