9.
Intrinsic error and
operating error
The intrinsic error is the maximum inaccuracy of the
FSA 050 under ideal conditions, whereas the operating
error is the maximum inaccuracy with allowance for
battery voltage, temperature, interference, system
voltage and system frequency.
Intrinsic error
Refer to Section 8 -Technical data.
Operating error
Function
Specification
Insulation range
±15 % ±2 positions
Continuity range
±26 % ±2 positions
Resistance range
±12 % ±2 positions
Voltage range
±10 % ±2 positions
Capacitance range
±18 % ±2 positions
Frequency range
±5 % ±2 positions