Exposure/Absorbed Dose Rates (photon) - Ion Chamber,
Energy Compensated GM, Tissue-Equivalent Plastic
Dose Equivalent Rates (neutron) - BF or He moderator,
Neutron-Proton Recoil (Rossi Detector, Liquid Plastic
Scintillator, Plastic/ZnS Scintillator) , LiGdBO -loaded Plastic
Beta and activity - Proportional Counter, GM, Plastic
Scintillator
Alpha activity - Proportional Counter, ZnS Scintillator, Air
Proportional, Solid-state Silicon, Plastic Scintillator
Alpha + beta activity - Proportional Counter, Plastic/ZnS
Scintillator, Plastic Scintillator, Solid-state Silicon
Gross gamma activity - NaI, CsI
X-ray spectroscopy - Si(Li)
Gamma spectroscopy - HPGe, CZT, HgI, CsI, LaBr
Alpha spectroscopy - Frisch Grid, Solid-state Silicon
Beta spectroscopy - BGO, Plastic Scintillator, Silicon
1.
Select an instrum ent appropriate for the isotope(s) to be surveyed
for.
2.
C heck instrum ent for a valid calibration sticker and for dam age that
would prevent it (them ) from operating acceptably.
3.
C heck the battery condition.
4.
P erform an operational (or perform ance) check.
5.
D eterm ine the isotope(s) correction factor to be applied to the
instrum ent.
6.
C alculate the instrum ent's M D A and com pare to survey criteria.
7.
If the instrum ent does not m eet all of the above criteria, then
replace the instrum ent (or change/charge the batteries) or change
your survey technique so that the instrum ent's M D A w ill m eet the
survey criteria.
9.
P erform and then docum ent the survey.

INSTRUMENT SELECTION AND USE