INSULIN PUMP
The Bluetooth 
®
word mark and logos are registered trademarks
owned by Bluetooth SIG, Inc. and any use of such marks by Roche
is under license. Other trademarks and trade names are those of
their respective owners.
User Guide
ACCU-CHEK, ACCU-CHEK SPIRIT COMBO, ACCU-CHEK AVIVA COMBO,
ACCU-CHEK FLEXLINK and ACCU-CHEK ULTRAFLEX are trademarks of Roche.
USA:
Manufactured for and distributed by:
Roche Insulin Delivery Systems Inc.
Fishers, IN 46037, USA
Toll free 1-800-280-7801
Manufacturer:
Roche Diagnostics GmbH
Sandhofer Strasse 116
68305 Mannheim, Germany
Swiss made
www.accu-chek.com

Advertisement

Table of Contents
loading

Summary of Contents for Accu-Chek Spirit Combo

  • Page 1 INSULIN PUMP The Bluetooth  ® word mark and logos are registered trademarks owned by Bluetooth SIG, Inc. and any use of such marks by Roche is under license. Other trademarks and trade names are those of their respective owners. User Guide ACCU-CHEK, ACCU-CHEK SPIRIT COMBO, ACCU-CHEK AVIVA COMBO, ACCU-CHEK FLEXLINK and ACCU-CHEK ULTRAFLEX are trademarks of Roche. USA: Manufactured for and distributed by: Roche Insulin Delivery Systems Inc. Fishers, IN 46037, USA Toll free 1-800-280-7801 Manufacturer: Roche Diagnostics GmbH Sandhofer Strasse 116 68305 Mannheim, Germany Swiss made www.accu-chek.com...
  • Page 2 Your insulin pump Your contact for assistance with and /or additional information about your Accu-Chek Spirit Combo insulin pump: Distributed in USA by: Roche Insulin Delivery Systems Inc. 11800 Exit 5 Parkway, Suite 120 Fishers, IN 46037, USA Toll Free 1-800-280-7801 www.accu-chek.com Menu key Up key OK key For technical support please contact Accu-Chek Customer Care at 1-800-688-4578 Down key Cartridge compartment with cartridge Adapter Luer lock of infusion set...
  • Page 3 Pump details and accessories Battery key Battery cover Battery Infrared window (next to Battery cover) Adapter Cartridge Compartment Cartridge Piston rod End plate Cartridge plunger...
  • Page 4 Overview of ADVANCED user menu Press d to move forward STOP Press d + a simultaneously to move backward Press f to select or save Press a to go up Press s to go down = Basal Rate Profile 2–4 BRP 4...
  • Page 5 Overview of STANDARD and CUSTOM user menus Accessible menus in CUSTOM user STOP menu can be selected using Accu-Chek PC software (sold separately). Press d to move forward Press d + a simultaneously to move backward Press f to select or save...
  • Page 6 (insulin glulisine, rDNA origin)** and ® NovoLog (insulin aspart)*** have been tested and found to ® be safe for use in the Accu-Chek Spirit Combo insulin pump. Humalog ® is a registered trademark of Eli Lilly and Company. ** Apidra ®...
  • Page 7 About this User Guide To ensure safe and convenient use of your new Accu-Chek Spirit Combo insulin pump, some information in this User Guide requires particular attention. This information is presented as “Warning”, “Caution” or “Note”. A “Warning” informs you about risks to your health. Neglecting this information may even lead to life-threatening situations in some cases.
  • Page 9: Table Of Contents

    Understanding your new system Your insulin pump and water Changing the infusion set Overview Changing the cartridge and infusion set Display and sounds General safety information Keys and key combinations Accessories and disposables Boluses and Temporary Basal Rates Emergency kit Understanding boluses General safety information Programming a Standard Bolus...
  • Page 10: Troubleshooting

    Using the Accu-Chek Spirit Combo insulin Appendix A: Technical Data pump with the Meter Appendix B: Configuration parameters Appendix C: Sterile products and accessories Meter Appendix D: Abbreviations Switching Bluetooth ON and OFF ® Appendix E: Beeps and melodies Pairing your pump with the Meter...
  • Page 11: Understanding Your New System

    Three user menus features that help to ensure safe and efficient pump therapy. Your Accu-Chek Spirit Combo insulin pump has been designed to adapt to your level of experience in pump therapy – whether you Note In this User Guide the term “Meter” always refers to the are a beginner or an experienced pump user.
  • Page 12 See chapter 5 Personalizing your pump. Data transfer Five basal rate profiles The infrared port on the bottom of your Accu-Chek Spirit Combo You can choose from five basal rate profiles, in line with your insulin pump enables you to transfer information between your different daily routines.
  • Page 13: Display And Sounds

    As well as Warnings and Errors, your pump can signal Reminders. This works like an electronic alarm clock. Your Accu-Chek Spirit Combo insulin pump has a Liquid Crystal Display (LCD) screen. This display, together with the four keys, Choice of batteries enables you to program the pump.
  • Page 14 When you reinsert the battery, the pump will perform a system check. If this does not resolve the problem, put your pump into STOP mode and contact your Accu-Chek Customer Care. For details see inside the back cover.
  • Page 15 See section 5.2 Selecting a user menu. Note To save battery life, the screen of your Accu-Chek Key Lock locked Spirit Combo insulin pump goes blank when you do not press any key for more than 20 seconds. As long as your Pump timer pump is showing a blank screen, you have to press any key before any other operation can be performed.
  • Page 16 The RUN screen also displays Reminder information (such as The STOP screen also displays Reminder information (such as “cartridge low”) and special functions (such as Key Lock – “battery low”) and special functions (such as Key Lock – locked locked or unlocked) in the form of symbols. (See Appendix F: or unlocked) in the form of symbols.
  • Page 17 1.2.4 Navigating through the screens By pressing f again you enter the BOLUS DATA function. You can view the last 30 boluses by pressing a or s. The following procedure explains how to navigate through the screens. Think of these screens as menu choices. See section 1.2.3 Display screens.
  • Page 18 In any sub-menu the same key combination will move you back Your Accu-Chek Spirit Combo insulin pump informs you of to the main menu screen. the status of your pump using audible beeps, vibrations and displayed messages.
  • Page 19: Keys And Key Combinations

    You then hear a melody, confirming that the STOP- Warning has been turned off. If you press the key too briefly, the You use the four keys on the Accu-Chek Spirit Combo insulin STOP-Warning will not be turned off.
  • Page 20 Turn off the STOP-Warning (press and tip of your fingernail, as this could damage the casing of your hold 3 sec. until you hear a melody) Accu-Chek Spirit Combo insulin pump. Instead, use the pad of your fingertip to avoid puncturing the keys.
  • Page 21 If the keys on your pump are not functioning properly, simultaneously. In the following table, a plus symbol (+) indicates disconnect your pump and contact your Accu-Chek Customer that both buttons must be pressed at the same time. Care (see inside back cover).
  • Page 22 The Key Lock feature enables you to lock all four keys on your sometimes be pressed several times until the desired value is Accu-Chek Spirit Combo insulin pump. This is a valuable safety shown, with each press of the key changing the figure by a small measure that prevents you from accidentally activating your increment.
  • Page 23: Accessories And Disposables

    1.4 Accessories and disposables WARNING High-quality sterile disposables and accessories are a vital Your pump has been designed for use with rapid-acting U100 part of the Accu-Chek Spirit Combo insulin pump system, and insulin. Humalog ® (insulin lispro, rDNA origin)*, Apidra ®...
  • Page 24 1.4.2 Adapter Note It is essential to change the adapter at least every two months. You must also change the adapter if the seals vents become worn, are lost, or if the vents are blocked. 1.4.3 Battery grip Your pump requires one AA 1.5 V battery in order to function. Use seal AA high-quality alkaline LR6 or lithium FR6 batteries that have a minimum capacity of 2500 mAh.
  • Page 25 Accu-Chek Spirit Combo insulin pump, see Appendix C. Note Always remember to remove the battery if your Accu-Chek Spirit Combo insulin pump is not going to be used for a long period, in order to conserve the battery. Battery cover The battery cover seals the battery compartment, and must be changed at least every two months.
  • Page 26: Emergency Kit

    Such an emergency kit should contain: Make sure that you have an alternative therapy plan, provided An Accu-Chek infusion set (see Appendix C) by your healthcare professional. Use this plan whenever you A new AA alkaline or lithium battery need to interrupt pump therapy.
  • Page 27: General Safety Information

    If you should drop your pump, immediately inspect it carefully to avoid inappropriate insulin delivery. This phrase means that for damage. Do not use your Accu-Chek Spirit Combo insulin either too much or too little insulin is delivered. Too much insulin pump if cracks or chips are visible –...
  • Page 29: Getting Started

    2.2 Inserting the battery This chapter explains how to prepare your pump and begin insulin pump therapy. Your Accu-Chek Spirit Combo insulin pump turns on as soon as you insert the battery. When you remove the battery, your pump 2.1 Before you begin keeps track of the time and date for about one hour.
  • Page 30 Procedure for inserting and changing the battery Remove or disconnect your infusion set from the infusion site, to ensure that there is no risk of accidental insulin delivery. Press d to move to the STOP YOUR PUMP screen. Press f to select. Your pump is now in STOP mode and insulin delivery stops.
  • Page 31 5. 5. 5. 6. 6. 6. Use the battery key to fully tighten the battery cover. You can verify that the battery cover is correctly tightened by checking that it is in line with the pump casing. Do not over-tighten. Place the cover on the positive end and gently push the battery in, while turning the cover clockwise to partially tighten it.
  • Page 32 CAUTION To prevent water from entering the pump casing, only insert or change the battery in a dry environment. Make sure that the battery cover seal is not worn or missing, and that the battery is correctly inserted. Note If your pump was in RUN mode before you removed the battery, the Error E8: POWER INTERRUPT will occur when the new battery is inserted.
  • Page 33: Start-Up Process

    Interruption of the performed. process may lead to malfunction of your pump. Start-up process sequence The software version screen appears. The Accu-Chek logo appears. The SELF-TEST screen appears.
  • Page 34 WARNING If you cannot hear the beeps from your Accu-Chek Spirit Combo insulin pump, you may not be notified of changes The BEEP TEST screen appears and your The VIBRATION TEST screen appears and within the pump in time to take action.
  • Page 35 The display turns black. Check that the If your pump detects a Warning or an After the self-test, press a or s to display is completely black. Error during the start-up process, it select the type of battery you inserted. signals the Error or Warning at this point.
  • Page 36: Setting The Time And Date

    2.4 Setting the time and date It is essential to set the time and date correctly, because basal WARNING rates and history information are stored on the basis of the time and date set in the pump. When you travel across time zones, Incorrect setting of the time and date can lead to incorrect and for the change to and from daylight savings time, make sure basal insulin delivery.
  • Page 37 Procedure for setting the time and date Press d to move to the TIME AND DATE The HOUR screen is displayed. Press d to move to the MINUTE screen. SETTINGS screen. Press a or s to set the hour. Press a or s to set the minute. Press f to select.
  • Page 38 Press d to move to the TIME FORMAT Press d to move to the DATE FORMAT screen. screen. Press a or s to set the time format. Press a or s to set the date format. Press f to save and exit.
  • Page 39: Your Basal Rate Profile

    2.5 Your basal rate profile Note For your safety and convenience during programming, Your Accu-Chek Spirit Combo insulin pump must be programmed with your personal settings before you begin pump therapy. Your your Accu-Chek Spirit Combo insulin pump switches to...
  • Page 40 Your basal rate profile should be reviewed: After any programming changes are made to your pump After a new battery is inserted After confirming the Error E7: ELECTRONIC ERROR. Note Insulin delivery does not stop or pause during basal rate programming. Procedure for programming your basal rate profile Press d to move to the BASAL RATE Press f to select.
  • Page 41 Note The increments shown depend on the programmed basal rate: it will 0.01 U from 0.05 U to 1.00 U, 0.05 U from 1.00 U to 10.0 U and Press a or s or scroll to set the 0.1 U from 10.0 U to 25.0 U Press d to move to the next hour.
  • Page 42 Copying an hourly basal rate If your basal rate is the same for multiple hours in a row, you can save time by copying an hourly basal rate to one or more hours. Press d to move to the BASAL RATE Press f to select.
  • Page 43 Review the new daily BASAL RATE TOTAL. undo the changes Wait for your Accu-Chek Spirit Combo insulin pump to return to the RUN or STOP screen, or Press f to save and exit. Press d and a simultaneously (exit feature) to exit the...
  • Page 44 Selecting a basal rate profile When you have programmed more than one basal rate profile (ADVANCED or CUSTOM MENU), you can select the required profile. Press d to move to the BASAL RATE Press f to select. SELECTION screen. The SELECT PROFILE for the current basal rate profile is displayed.
  • Page 45 Note When you change your user menu for example from Basal rate and bolus ADVANCED to STANDARD, your current basal rate profile During normal use, your pump is in RUN mode and insulin is may not appear. continuously delivered to your body, according to the basal rate profile you have programmed.
  • Page 46: Connecting The Cartridge, Adapter And Infusion Set

    Before you change the cartridge, adapter or infusion set, Have the following materials ready: check that your pump is in good working order. A filled Accu-Chek Spirit 3.15 ml Cartridge System with a See section 7.1.1 Checklist. luer-lock connection (for filling instructions see the instructions...
  • Page 47 Procedure to connect the cartridge, adapter and infusion set Push the adapter all the way onto the Carefully prepare a new Accu-Chek Remove the protective cap from the filled cartridge tip, to the stop. infusion set, following the instructions cartridge. (Do not touch the cartridge tip.) supplied for use.
  • Page 48: Inserting The Cartridge

    (insulin aspart)*** have been tested and found to A new Accu-Chek infusion set be safe for use in the Accu-Chek Spirit Combo insulin pump. Before using a different insulin with the Accu-Chek Spirit Combo insulin pump, you should check the insulin label to make sure it can be used with this device.
  • Page 49 Press d to move to the CARTRIDGE Make sure your infusion set is discon- CHANGE screen. nected and the cartridge has been Hold the pump with the adapter facing removed. the ground. Remove the cartridge, if Press f to select. necessary.
  • Page 50 Hold your pump upright. Hold the new see step 10 filled cartridge so that the connected adapter and infusion set points upwards, parallel, and close to the cartridge compartment. see step 11 Make sure that the top edge of the threaded part of the adapter is level with the top of the cartridge compartment.
  • Page 51 It is not necessary to apply pressure. Do not over-tighten the adapter, other- wise it will be difficult to remove the cartridge, adapter and cartridge plunger, and could cause damage to your pump. Position your insulin pump upright (with the adapter pointing upwards). Insert the cartridge into the cartridge compartment.
  • Page 52 See section 2.8 Priming the infusion set. Start from step 4 in the Procedure for priming the infusion set. Press f after correctly inserting the When the self-test process is finished, cartridge. the INFUSION SET PRIMING screen is displayed. Your pump performs a self-test. See section 2.3 Start-up process.
  • Page 53: Priming The Infusion Set

    The Accu-Chek Spirit Combo insulin pump stops priming at body. 25 units of insulin. If no insulin emerges from the needle or there...
  • Page 54 Procedure for priming the infusion set Press d to move to the STOP YOUR The pump displays the STOP screen, and PUMP screen. insulin delivery stops. Prepare the tubing or cannula for priming Press f to select. according to the infusion set’s instruc- tions for use.
  • Page 55 The infusion set is properly primed when there are no air bubbles visible in the tubing and a drop of insulin has come out of the needle. When the priming process is complete, your pump returns to STOP. A count-up is shown on the display. Your pump stops priming after 25 units.
  • Page 56: Positioning The Infusion Set

    Always follow the instructions given to you by your doctor or Have the following materials ready: healthcare team and the instructions for the infusion set you Your Accu-Chek Spirit Combo insulin pump are using. A disinfectant for your skin A new Accu-Chek infusion set Change your infusion set as frequently as recommended by the manufacturer.
  • Page 57 Make sure your pump is in STOP mode. Wash your hands thoroughly. Disinfect the infusion site and let it dry Insert the cannula of your Accu-Chek completely. infusion set according to the instructions.
  • Page 58 Accu-Chek Spirit Combo set you are using. Failure to do so could result in missed insulin pump.
  • Page 59: Starting Insulin Delivery

    2.10 Starting insulin delivery Procedure for starting insulin delivery Your pump is in STOP mode. Press d to move to the START YOUR Press f to select. Your pump shows the PUMP screen. RUN screen. Insulin delivery begins within 3 minutes.
  • Page 60: Stopping Insulin Delivery

    CAUTION actions when the pump is in STOP mode. These functions and actions include: When your Accu-Chek Spirit Combo insulin pump is in STOP changing the cartridge mode, no insulin is delivered to your body. You must put your connecting and disconnecting the adapter and / or infusion set pump into RUN mode to continue insulin delivery.
  • Page 61 Do not re-use single-use materials because this might lead to seconds. You then hear a melody and / or feel a vibration, a malfunction of your Accu-Chek Spirit Combo insulin pump, confirming that the STOP-Warning has been turned off. incorrect insulin delivery and / or infection. Always handle these items with clean hands.
  • Page 62 The support of a doctor or healthcare team is vital when pump, in order to avoid incorrect insulin delivery. Record your training on and using your Accu-Chek Spirit Combo insulin personal settings from your current insulin pump, and ensure pump. Regular visits to your doctor or healthcare team are also that they are correctly programmed into your new Accu-Chek essential during insulin pump therapy.
  • Page 63 Note Your pump cannot detect infusion set leakage. If you detect any loss of insulin and all parts are correctly tightened, immediately replace the leaking component. Because insulin delivery has been interrupted, check your blood glucose level and take appropriate action according to your doctor’s or healthcare team’s instructions.
  • Page 65: Using Your Pump In Daily Life

    Using your pump in daily life 3.1.1 When not to wear your pump This chapter provides important guidelines on how to use your pump safely from day to day. Electromagnetic fields and hazardous areas 3.1 Wearing your pump WARNING Attach your pump safely to your body or clothes to prevent Do not use your pump near electromagnetic fields such as damage.
  • Page 66 Exercise is an important part of diabetes management. Your department stores, should not normally affect the functioning of Accu-Chek Spirit Combo insulin pump is designed to enable you your pump. to safely perform a range of sporting activities while wearing your pump.
  • Page 67 If you have questions, contact your Accu-Chek Customer Care and tubing and also measure your blood sugar more frequently in such situations.
  • Page 68: Interruption Of Insulin Pump Therapy

    3.2 Interruption of insulin pump therapy Consult with your doctor or healthcare team about when and for how long your insulin pump therapy can be interrupted. Measure your blood glucose level regularly during any insulin delivery interruption. Use a syringe or insulin pen to inject insulin according to the instructions of your doctor or healthcare team.
  • Page 69 Resuming insulin delivery Where appropriate, attach a new Accu-Chek infusion set. (See section 3.4 Changing the infusion set.) Measure your blood glucose level within two hours of restarting pump therapy to...
  • Page 70 3.2.2 Long interruptions When you need to interrupt your pump therapy for more than one day, use the following procedure. Put your pump into STOP mode. Remove the cartridge, adapter, and Remove the battery and reinsert the infusion set. battery cover. Store your pump properly.
  • Page 71: Your Insulin Pump And Water

    3.3 Your insulin pump and water 3.3.1 Accidental contact with water CAUTION Your Accu-Chek Spirit Combo insulin pump is protected against Before any contact with water, disconnect and take off your Accu-Chek Spirit Combo insulin pump. Check your pump daily brief accidental contact with water.
  • Page 72 CAUTION Put your pump into STOP mode, disconnect the pump from your Avoid any contact between your Accu-Chek Spirit Combo body, and inspect it. Use a soft cloth to dry the outside of the insulin pump, the infusion set, and the connecting parts of...
  • Page 73: Changing The Infusion Set

    3.4 Changing the infusion set When you change the infusion set, keep the following information Always disconnect your infusion set before changing a in mind: cartridge or priming the infusion set. Otherwise, you risk Infusion sets are sterile products intended for single use only. uncontrolled insulin delivery into your body.
  • Page 74 Procedure for changing the infusion set Have the following materials ready: Your Accu-Chek Spirit Combo insulin pump A disinfectant for your skin A new Accu-Chek infusion set Make sure your pump is in STOP mode. Wash your hands thoroughly. 5. 5.
  • Page 75 8. 8. 8. Screw the Accu-Chek infusion set Remove the used infusion headset and Carefully prepare a new Accu-Chek clockwise into the adapter. dispose of it correctly. infusion set for use. Carefully tighten the luer-lock connector by hand. Prime the infusion set following the Put the pump in RUN mode.
  • Page 76 Note If you use a soft-cannula type infusion set, you must begin with a bolus to fill the air space in the cannula and headset, according to the instructions for the infusion set you are using. Failure to do so could result in inadequate insulin dosing.
  • Page 77: Changing The Cartridge And Infusion Set

    Your Accu-Chek Spirit Combo insulin pump A disinfectant for your skin A new Accu-Chek infusion set A filled Accu-Chek Spirit 3.15 ml Cartridge System A new adapter (this must be changed at least every two months) Procedure for changing the cartridge and infusion set Make sure your pump is in STOP mode.
  • Page 78 If this should happen, please call your Accu-Chek Customer Care (see inside back cover). Hold your Accu-Chek Spirit Combo insulin Remove the adapter and cartridge from pump with the adapter pointing down- your pump by unscrewing the adapter ward.
  • Page 79 Hold the adapter to the light to check for If necessary, clean the adapter using Insert a new cartridge and a new infusion signs of wear or dirt (especially on the water and dry it. If it shows signs of wear set.
  • Page 80 (insulin glulisine, rDNA ® origin)** and NovoLog ® (insulin aspart)*** have been tested and found to be safe for use in the Accu-Chek Spirit Combo insulin pump. Humalog is a registered trademark of Eli Lilly and Company. ® ** Apidra is a registered trademark of the sanofi-aventis group.
  • Page 81: Boluses And Temporary Basal Rates

    Boluses and Temporary Basal Rates 4.1 Understanding boluses Your pump can deliver three types of bolus: A bolus is the amount of insulin delivered (in addition to the basal Standard Bolus fast delivery rate) to cover the intake of food and to correct high blood glucose levels.
  • Page 82: Programming A Standard Bolus

    4.2 Programming a Standard Bolus There are two ways to program a Standard Bolus: Quick Bolus Using a and s on your insulin pump, with bolus increments determined by the user Using Standard Using d and f to enter the menu Bolus menu and then a and s to program the amount...
  • Page 83 4.2.1 Programming a Quick Bolus The following process uses only the a and s keys. You can looking at the pump screen. After the bolus is programmed, you start with either key. The recommended steps below are those can count the beeps and vibrations of your pump to check the described the Down key s in step 1 and the Up key a in Step correct programming.
  • Page 84 WARNING The bolus increment that can be programmed into your Accu-Chek Spirit Combo insulin pump determines the Quick Bolus amount using the a and s keys of your insulin pump. Incorrect programming of the bolus increment may cause inappropriate insulin delivery.
  • Page 85 4.2.2 Cancelling a Quick Bolus During programming Press the s or a key you used to enter the Quick Bolus programming No bolus is delivered. (the bolus amount option. This will reset the bolus amount to 0.0 units. You hear a melody. blinks): If no new bolus is programmed, your pump returns to the RUN screen and beeps three times and vibrates after five seconds.
  • Page 86 4.2.3 Programming a Standard Bolus using Standard bolus menu Independent of the bolus increment setting, the bolus increments for the menu-guided Standard Bolus are fixed at 0.1 unit. You can program this bolus in the STANDARD BOLUS screen by holding the a or s key (scrolling) until the bolus amount is Press d to move to the STANDARD displayed.
  • Page 87 After this bolus delivery start delay, your pump beeps three times and vibrates and begins to deliver the total bolus amount programmed. The countdown of the remaining bolus amount appears on the display. 4.2.4 Cancelling a Standard Bolus During programming You have three options for cancelling during programming: No bolus is delivered.
  • Page 88: Programming An Extended Bolus

    4.3 Programming an Extended Bolus Cancellation during the start delay and during the bolus delivery will cause a WARNING W8: BOLUS CANCELED. Press f twice to confirm and snooze the Warning. See chapter 8. The Extended Bolus option enables you to program a bolus to be delivered over a period of time.
  • Page 89 Tip: You can program a Standard or Quick Bolus while an If you want to program a different Extended Bolus, put the pump Extended Bolus is ongoing. If you cancel the Standard Bolus, the in STOP mode to cancel the current bolus. Then put the pump in Extended Bolus continues to run.
  • Page 90 Tip: Press d to switch between programming the bolus amount and bolus duration. Press d to move to the BOLUS Press f to confirm the bolus amount DURATION screen. and duration. The pump returns to the RUN screen. The bolus delivery begins Press a to increase or s to decrease within the next 3 minutes.
  • Page 91 4.3.1 Cancelling an Extended Bolus During programming You can either: No bolus is delivered. (the bolus amount or wait for your pump to return to the RUN screen. bolus duration blinks): press d and a simultaneously to exit the Function screen. set the bolus amount to 0.0 units and press f.
  • Page 92 Note If a Temporary Basal Rate is active at the same time, See chapter 8 Warning W6: TBR CANCELLED and it is also cancelled when your pump is put into Warning W8: BOLUS CANCELLED. Make sure that the STOP mode. A Warning W8: BOLUS CANCELLED and cancellation was deliberate and then program a new a Warning W6: TBR CANCELLED are displayed.
  • Page 93: Programming A Multiwave Bolus

    4.4 Programming a Multiwave Bolus The Multiwave Bolus option is available in the ADVANCED user When programming a Multiwave Bolus, make sure that you are menu. It is designed to better simulate the body’s insulin using the ADVANCED user menu and your pump is in RUN mode. delivery, and combines an immediate bolus delivery followed by an Extended Bolus delivery.
  • Page 94 Procedure for programming a Multiwave Bolus Press d to move to the Multiwave Bolus Press f to select. The BOLUS AMOUNT Press a or s to increase or decrease screen. screen appears. the total bolus amount. Note The first time you press a, the bolus duration of the last Multiwave Bolus and a default immediate bolus amount appear.
  • Page 95 Tip: Press d to loop between program- ming the total BOLUS AMOUNT, IMMEDI- ATE BOLUS and BOLUS DURATION. Check the total and immediate bolus Press f to confirm both bolus amounts amounts and the Extended Bolus duration and the bolus duration. The Multiwave programmed on the screens.
  • Page 96 4.4.1 Cancelling a Multiwave Bolus During programming There are three ways to cancel a Multiwave Bolus during programming: No bolus is delivered. (the bolus amount Do not press any key for 20 seconds. The pump returns to the RUN screen. blinks): Return to the Multiwave Bolus screen by pressing d + a simultaneously.
  • Page 97 4.5 Programming a Temporary Note If a Temporary Basal Rate is active at the same time, it is Basal Rate (TBR) also cancelled when your pump is put in STOP mode. The Warning W8: BOLUS CANCELLED and a Warning W6: TBR CANCELLED is displayed.
  • Page 98 Note A running Temporary Basal Rate increase or decrease remains active, even if you have changed the basal rate profile. You can only program a Temporary Basal Rate when your pump is in RUN mode. Procedure for programming a Temporary Basal Rate Note If no Temporary Basal Rate is currently active, the percentage is set to 100 %.
  • Page 99 Tip: When you increase (or decrease) the Temporary Basal Rate, the duration of your previous Temporary Basal Rate increase (or decrease) appears. However, Press a or s to increase or decrease when you use your pump for the first Press d to move to the TBR DURATION the basal rate.
  • Page 100 Note During Temporary Basal Rate delivery, the increase or TBR OVER Alert decrease in percentage, the increased or decreased At the end of a Temporary Basal Rate, a Warning W7: TBR OVER hourly basal rate (hourly basal rate plus TBR percentage), is displayed.
  • Page 101 4.5.1 Cancelling a Temporary Basal Rate During programming: There are three ways to cancel a Temporary Basal Rate during programming: Pump continues to Do not press any key for 20 seconds. The pump returns to the RUN screen. deliver normal basal Exit the Temporary Basal Rate (TBR) screen by pressing a + d rate.
  • Page 102 4.6 General safety information Note If an Extended Bolus or a Multiwave Bolus is active at the WARNING same time, it is also cancelled when your pump is put into STOP mode. A Warning W8: BOLUS CANCELLED and a Incorrect programming of your insulin pump may cause Warning W6: TBR CANCELLED are displayed.
  • Page 103: Personalizing Your Pump

    Your Accu-Chek Spirit Combo insulin pump has been designed to adapt to your level of experience in pump therapy – whether you are a beginner or an experienced pump user.
  • Page 104 Note For your safety and convenience during programming, become more experienced, you may wish your Accu-Chek Spirit Combo insulin pump returns to the to change to the ADVANCED user menu. RUN or STOP screen if no key is pressed within 20 seconds.
  • Page 105: Selecting A User Menu

    5.2 Selecting a user menu Note For guidance in choosing the most suitable user menu, see section 5.1.1 User menus. If you are still not sure which menu to choose, discuss this with your doctor or healthcare team. Procedure for selecting a user menu Press d to move to the MENU SETTINGS The SELECT USER MENU screen is Press f to confirm.
  • Page 106 … if the previous basal rate profile is not Press a or s to select a basal rate available from the selected user menu, profile, if required. dashes appear instead. Check the (new) basal rate profile and its daily insulin total. Press f to save and exit.
  • Page 107: Standard User Menu

    5.3 STANDARD user menu The following is a list of menu options from the STANDARD user menu. To view the order in which they appear on the display, refer to the menu diagram below: Press d to move forward STOP Press d + a simultaneously to move backward...
  • Page 108: Programming A Temporary Basal Rate (Tbr)

    See section 5.2 Selecting a user menu earlier in this chapter. INFUSION SET PRIMING See section 2.8 Priming the infusion set. BLUETOOTH SETTINGS See chapter 6 Using the Accu-Chek Spirit Combo insulin pump START YOUR PUMP with the Meter. See section 2.10 Starting insulin delivery.
  • Page 109: Viewing Your Pump Data

    5.4 Viewing your pump data The memory of your Accu-Chek Spirit Combo insulin pump stores Bolus history last 30 boluses all the events (Warnings and Errors, programming operations and insulin delivery records) up to 4500 events. This data corresponds Alarm history...
  • Page 110 5.4.1 Viewing your bolus history Each BOLUS DATA screen displays: Bolus type (Standard, Extended or Multiwave Bolus) The BOLUS DATA screen allows you to view up to the last Bolus duration (only displayed if an Extended or 30 bolus deliveries, starting with the most recent entry. Multiwave Bolus) Time Date...
  • Page 111 5.4.2 Viewing the Error and Warning history The alarm data menu allows you to view up to the last Each alarm data screen displays: 30 Warnings and Errors, starting with the most recent entry. Number and type of the Warning or Error (for example, W6) Warning or Error (for example, TBR CANCELLED) Time Date...
  • Page 112 5.4.3 Viewing the history of daily insulin totals This function allows you to view up to the last 30 daily totals of Each screen displays: insulin delivered (from midnight to midnight; including basal rate Daily total of insulin delivered plus boluses), starting with the most recent entry. Date Entry number (01 is the most recent) of total entries (example 01/04) of a daily insulin total...
  • Page 113 5.4.4 Viewing the Temporary Basal Rate history This option allows you to view up to the last 30 Temporary Basal Each TBR DATA screen displays: Rate (TBR) increases or decreases, starting with the most recent TBR increase or decrease in percentage entry.
  • Page 114 5.4.5 Viewing the time remaining (pump timer) symbol appears in the RUN screen or STOP screen. When the timer expires, an Error occurs, and your pump goes into STOP For loan pumps and “Backup” pumps, a pump timer limits the mode.
  • Page 115: Transferring Your Data To A Computer

    5.5 Transferring your data to a computer The built-in infrared interface on the bottom of your Accu-Chek information management products (for details see the user Spirit Combo insulin pump enables you to transfer data between information provided with the software). your pump and a PC.
  • Page 116 Set up your pump with a cartridge, adapter, and a new infusion set, and put it in RUN mode if necessary. Note Your pump will return to STOP mode 15 minutes after the last data transfer when the DATA TRANSFER screen is When the data transfer is complete, selected.
  • Page 117: Pump Settings

    The Key Lock function allows you to lock the four keys of your settings relating to your therapy, see 5.10 Therapy settings later Accu-Chek Spirit Combo insulin pump. This is an important in this chapter. safety measure that prevents you from accidentally activating functions, for example while sleeping or during contact sports.
  • Page 118 Procedure for unlocking the Key Lock If you press any key while the screen is blank, the Key Lock symbol indicates that the keys are locked. Note The Key Lock remains unlocked during fast bolus delivery of Standard or Multiwave Bolus. Key Lock is unlocked.
  • Page 119 5.6.2 Adjusting the sound volume Note The volume of the STOP-Warning is independent of the Your pump emits different acoustic signals in different handling situations to facilitate handling e.g. for blind or partially-sighted programmed beep volume – it always sounds at the users.
  • Page 120 5.6.3 Changing signal settings Beep only The Accu-Chek Spirit Combo insulin pump enables you to choose Vibrate only how you want your pump to alert you when a Warning or Error occurs. You have three options: Beep and vibrate For your safety, if you do not respond to the vibration, the pump starts beeping after one minute.
  • Page 121 5.6.4 Display orientation Note If you turn the display orientation of your pump by 180°, the functions of the a and s keys are also reversed. Your pump allows you to turn the display by 180°, so that you The functions of the d and f keys remains the same, can view it more easily, depending on how you are wearing the whichever screen orientation you choose.
  • Page 122 5.6.5 Setting the display contrast Your pump allows you to adjust the display contrast to suit your preferences. Procedure for setting the display contrast Press d to move to the PUMP Press d to move to the DISPLAY SETTINGS screen. CONTRAST screen.
  • Page 123 5.6.6 Selecting the language Your pump can display the information in several languages. You can chose the language according to your preferences. Procedure for setting the language Press d to move to the PUMP Press d to move to the LANGUAGE SETTINGS screen.
  • Page 124: Advanced User Menu

    5.7 ADVANCED user menu The following is a list of additional menu options available in the ADVANCED user menu. To view the order in which they appear on the display, refer to the menu diagram below: Press d to move forwards STOP Press d + a simultaneously to move backwards...
  • Page 125: Basal Rate Profiles

    5.8.1 Programming additional basal rate profiles WARNING Your Accu-Chek Spirit Combo insulin pump must be pro- grammed with your personal settings prior to starting insulin pump therapy. Do not operate your pump without knowing your personal settings. If you are not sure about your personal settings or you are inexperienced, ask your doctor or healthcare team to check your personal settings.
  • Page 126 Procedure for Programming additional basal rate profiles See section 2.5.1 Programming your basal rate profile. Press d to move to the BASAL RATE The hourly basal rates for each additional PROGRAMMING (1, 2, 3, 4, 5) screen you basal rate profile are set in the same way wish to program.
  • Page 127 5.8.2 Selecting a basal rate profile Procedure for selecting a basal rate profile Press d to move to the BASAL RATE Press a or s to select a basal rate SELECTION screen. profile. The selected basal rate profile and its daily insulin total appear. Press f to enter.
  • Page 128: Setting A Reminder

    5.9 Setting a Reminder You can set both single and multiple Reminders on your pump. Multiple Reminders are repeated at a set time every day. You can use this feature to remind you, for example, when you need to test your blood glucose. Procedure for setting single or multiple Reminders Press d to move to the REMINDER Press f to select.
  • Page 129 Press d to move to the HOUR screen. Press d to move to the MINUTE screen. Press a or s to set the hour. Press a or s to set the minute. Press f to save and exit. Turning off the Reminder When the Reminder sounds, the Reminder screen is displayed.
  • Page 130: Therapy Settings

    5.10 Therapy settings These settings relate to the therapeutic use of the pump. For your Accu-Chek Spirit Combo insulin pump is adjustable. It is settings relating to the practical use of the pump, see Pump initially set to 0.5 units per key-press, but it can be changed settings earlier in this chapter.
  • Page 131 5.10.2 Adjusting the prime amount Note The amount of insulin used for the priming is not added to The prime amount needed to fill an infusion set depends on the length of the infusion set tubing. The shorter your infusion set the history of daily insulin totals.
  • Page 132 5.10.3 Locking basal rate profiles Your Accu-Chek Spirit Combo insulin pump allows you to adapt If you try to program a basal rate while the BASAL RATE LOCK is your hourly basal rate profile as needed. You can then lock...
  • Page 133 5.10.4 Automatic Off Automatic Off is a safety feature that stops insulin delivery and triggers Error E3: AUTOMATIC OFF if no keys are pressed within a programmed time period in RUN mode. This feature can either be set to OFF or programmed up to 24 hours in 1-hour intervals. The factory setting is OFF.
  • Page 134: Custom User Menu

    5.11 CUSTOM user menu The CUSTOM user menu is designed to adapt your pump to your special requirements and preferences. It allows you or your doctor or healthcare team to display or hide screens and thereby customize the individual user menus. STOP Press d to move forward Press d + a simultaneously...
  • Page 135 Note By default, the CUSTOM user menu contains all of the functions available except the following: Multiwave Bolus Basal rate profiles 3, 4 and 5 Therapy Settings To change the CUSTOM user menu, you, your doctor, or your healthcare team must use the appropriate Accu-Chek PC software.
  • Page 137: Pump With The Meter

    Using the Accu-Chek Spirit Combo insulin pump with the Meter 6.1 Meter This chapter describes how the Accu-Chek Spirit Combo insulin pump works with the Meter, and explains how to pair the two devices using Bluetooth ® wireless technology. The Meter is both a blood glucose meter and a remote-control device for your Accu-Chek Spirit Combo insulin pump.
  • Page 138 ® settings Display settings All the other functions available on the Accu-Chek Spirit Combo insulin pump can be performed using the Meter. Before using your pump together with the Meter you must read the documentation (Owner’s Booklet) provided with this product.
  • Page 139: Switching Bluetooth

    6.2 Switching Bluetooth ON and OFF ® To prolong the battery life and avoid accidental changes on your WARNING pump, you should switch off Bluetooth ® wireless technology, if the Meter is lost, stolen or, you plan not to carry it with you for You must ensure that Bluetooth ®...
  • Page 140: Pairing Your Pump With The Meter

    6.3 Pairing your pump with the Meter Your Accu-Chek Spirit Combo insulin pump and your Meter will You should not perform the pairing process in crowded areas, already have been paired by default. If you need to replace one since the signals of other electronic devices could interfere with of the devices for any reason you need to redo the pairing.
  • Page 141 Note When Bluetooth is turned off, ® you cannot access the Paired If the pump screen indicates that no Device menu. paired device is available, turn on the Meter in pairing mode by holding down Pump screen * and simultaneously pressing +. Meter screen The pump begins searching and detects the Meter, displaying its serial number...
  • Page 142 The pump begins the pairing process and shows a 10-digit code. Enter the code on the Meter display, using =, Pump screen and then to move to the next digit. Confirm the complete code with .. Meter screen After the pairing is completed press + to turn off the Meter.
  • Page 143: Troubleshooting The Bluetooth ® Connection

    6.4 Troubleshooting the Bluetooth ® connection In case of problems with the Bluetooth ® connection or after the appearance of the W10: BLUETOOTH FAULT, you have to repeat the pairing procedure described in section 6.3 Pairing your pump with the Meter. Since only one Meter at a time can be paired with your pump, you must delete the pairing before you can pair your pump with a new Meter.
  • Page 144 6.4.1 Handling Warnings and Errors The Accu-Chek Spirit Combo insulin pump and Meter behave in the following ways when Warnings and Errors occur. If both devices are switched on and paired you can enter the Meter’s Pump menu and snooze and confirm the Warning or Press d to move to the DELETE DEVICE Error through the Meter.
  • Page 145 6.4.2 Use in aircraft WARNING You must ensure that Bluetooth ® is switched off on both devices before boarding an aircraft, as emissions from your pump or the Meter could interfere with the electronics systems of the aircraft.
  • Page 147: Care And Maintenance

    Spirit Combo insulin pump as this may void the guarantee. For further information, contact your Accu-Chek Customer Care (see inside back cover).
  • Page 148 7.1.1 Checklist Check your pump daily, and make sure that: Your infusion site is secure, comfortable, and free of irritation The pump casing, the display, and the cartridge are not or infection. chipped or cracked and that the display does not show The cartridge is free of air bubbles.
  • Page 149: Damage To Your Pump

    Press f twice to confirm and turn off the alert. The Accu-Chek Customer Care (details on inside back cover). alerts are recorded in the alarm history. If you drop your pump:...
  • Page 150 Avoid contact with water. Check daily that your Accu-Chek your Accu-Chek Spirit Combo insulin pump if chips, cracks Spirit Combo insulin pump is not chipped, cracked or damaged or other damages are visible. Otherwise water, dust, insulin in any way, and that the battery cover and the adapter are or other foreign substances may enter your pump and could correctly closed.
  • Page 151 Other liquids You do not have to worry about contact between your pump and Your Accu-Chek Spirit Combo insulin pump is IPX8 waterproof perspiration or saliva. You must however check your pump when leaving the factory site. However hair-line cracks could immediately after contact with other liquids or chemicals such as: appear during use.
  • Page 152: Repair Of Your Pump

    Accu-Chek Customer Care. By returning your pump to Roche Insulin Delivery Systems Inc., If you have problems with your pump, contact your Accu-Chek you authorize Roche Insulin Delivery Systems Inc. to undertake Customer Care (details on inside back cover). Your Accu-Chek...
  • Page 153: Maintaining Your Pump

    The life of any battery is affected by how you necessary, use a clean cloth with alcohol (isopropanol or use your Accu-Chek Spirit Combo insulin pump, your personal ethanol). If these measures are not successful, contact your settings, delivery rates, temperature, and other factors.
  • Page 154: Storing Your Pump

    CAUTION contact with blood during use, it may carry a risk of infection. If you need to dispose of your Accu-Chek Spirit Combo insulin When your pump will not be in use for a long period of time, it must be properly stored to prevent any subsequent malfunction.
  • Page 155: Warnings And Errors

    A safety system monitors and controls the function of your If you do not confirm a Warning or Error within 60 seconds, it will Accu-Chek Spirit Combo insulin pump. This safety system be signalled again. Even when the beep is turned off or pro- performs more than 9 million safety checks every day.
  • Page 156: Confirming A Warning Or Error

    8.1 Confirming a Warning or Error When you confirm: When a Warning or an Error occurs: W1: CARTRIDGE LOW Press f to turn off the beeps and vibrations. The Warning or E1: CARTRIDGE EMPTY Error code stays on the display. W2: BATTERY LOW E2: BATTERY EMPTY W5: PUMP TIMER (“Backup”...
  • Page 157: List Of Warnings

    In order to maintain insulin delivery, you must act immediately according to the instructions given for each Error code, or put your Accu-Chek Spirit Combo insulin pump in RUN mode to continue the insulin delivery. If necessary, put your pump in RUN mode.
  • Page 158 Warning W2: BATTERY LOW Warning W3: REVIEW TIME The battery power is low. Be prepared for a battery change as The pump has been without power for more than one hour. The soon as possible. The battery low symbol appears in the RUN or pump settings and history data are saved in memory regardless STOP screen as a Reminder until you change the battery.
  • Page 159 The pump has reached six years of life. You may want to Your pump’s operating time will expire soon. When the Warning consider updating to a new product. Please contact Accu-Chek W5: PUMP TIMER occurs, the pump will display the symbol to Customer Care (see inside back cover for phone number).
  • Page 160 STOP mode and will no longer operate. An Error, E5: END OF OPERATION, will appear on the display. Contact your Accu-Chek Customer Care (details on inside back cover) to discuss the options for further use of your Accu-Chek Spirit Combo insulin If necessary, put your pump in RUN mode. pump.
  • Page 161 Warning W7: TBR OVER (Temporary Basal Rate Over) Warning W8: BOLUS CANCELLED A Temporary Basal Rate has ended. The original basal rate A bolus was cancelled during the start delay or after delivery (100 %) will resume automatically once the Warning has began.
  • Page 162 Warning W10: BLUETOOTH FAULT The pump is coming to the end of its loan period. In the MY DATA You need to re-do the pairing procedure for the Accu-Chek menu, check the remaining loan time in the PUMP TIMER menu.
  • Page 163: List Of Errors

    Reconnect the infusion set or insert a new one. WARNING Put your pump in RUN mode. After the occurrence of an Error, your Accu-Chek Spirit Combo insulin pump is in STOP mode and insulin delivery is inter- See section 3.5 Changing the cartridge and infusion set.
  • Page 164 Press f twice to snooze and to confirm the Error. Error E3: AUTOMATIC OFF Automatic Off is a safety feature that stops insulin delivery and triggers an Error E3: AUTOMATIC OFF if no keys are pressed within a programmed time period in RUN mode. Press f twice to snooze and to confirm the Error.
  • Page 165 Error E4: OCCLUSION If an Error E4: OCCLUSION (blockage) occurs, follow the steps listed below: WARNING Press f twice to turn off and to confirm the Error. If the Error E4: OCCLUSION occurs, immediately check your blood glucose level, and then check again about two hours later, as insulin delivery has been interrupted.
  • Page 166 Follow the steps listed below: healthcare team. Contact your Accu-Chek Customer Care (details on inside back cover) for more assistance. Press f twice to turn off and to confirm the Error.
  • Page 167 Press f twice to snooze and to confirm the Error. Contact your Accu-Chek Customer Care (see inside back cover) to discuss the options for further use of the Accu-Chek Spirit Combo insulin pump.
  • Page 168 If the Error E6: MECHANICAL ERROR occurs again, use the Press f twice to snooze and to confirm the Error. alternative therapy plan provided by your doctor or healthcare team. Contact your Accu-Chek Customer Care (details on inside back cover) for more assistance. Error E7: ELECTRONIC ERROR An electronic Error has been discovered during a routine Remove or disconnect the infusion set from your infusion site.
  • Page 169 If the Error E7: ELECTRONIC ERROR occurs again, contact your Check the time and date and correct them if they are wrong. Accu-Chek Customer Care (details on inside back cover) for See section 2.4 Setting the time and date. assistance. Use the alternative therapy plan provided by your Put your pump in RUN mode.
  • Page 170 Press f twice to snooze and to confirm the Error. Contact your Accu-Chek Customer Care (details on inside back cover) to discuss the options for further use of the Accu-Chek Remove or disconnect the infusion set from your infusion site.
  • Page 171 Prime the infusion set. See sections 3.5 Changing the cartridge and infusion set and Reconnect the infusion set at your infusion site. 2.8 Priming the infusion set. Put your pump into RUN mode. If the Error E10: CARTRIDGE ERROR occurs during the WARNING rewinding of the piston rod, insert a new battery in the pump and repeat the procedure.
  • Page 172 Error E12: DATA INTERRUPTED will occur. The data transfer must be successfully completed before you Restart the data transfer. can put your Accu-Chek Spirit Combo insulin pump into RUN mode. Before you can put your pump in RUN mode, the data transfer must have been successfully performed.
  • Page 173 Error E13: LANGUAGE ERROR The language setting needs to be checked. Press f twice to snooze and to confirm the Error. Follow the steps in section 5.6 Pump settings to change the language. Put your pump in RUN mode.
  • Page 174: List Of Reminders

    8.4 List of Reminders Reminder R1: Reminder A Reminder has been programmed on the pump. Press f twice to snooze and to confirm the Reminder. If necessary, put your pump in RUN mode. Note If you have set the Reminder occurrence to EVERY DAY, the same Reminder occurs every 24 hours.
  • Page 175: Troubleshooting

    Note Troubleshooting information relating to the Bluetooth ® connection between the Meter and the Accu-Chek Spirit Combo insulin pump is included in chapter 6 Using the Accu-Chek Spirit Combo insulin pump with the Meter. Also refer to the documentation (Owner’s Booklet) that...
  • Page 176: Therapy-Related Issues

    Problem Recommended Action Individual characters, numbers, or symbols on the display Contact your Accu-Chek Customer Care (details on inside back cover). are shown incompletely or not at all. Your Accu-Chek Spirit Combo insulin pump does not beep Check that the beep volume and alarm signals are turned on. Check the or vibrate.
  • Page 177 Problem Recommended Action There are air bubbles in the cartridge and /or infusion set. Check your blood glucose level. In the cartridge: Disconnect the infusion set from the headset and re-prime the infusion set. If the air bubbles remain, replace the cartridge.
  • Page 178 Problem Recommended Action There is a leak in the system. Check your blood glucose level. Check for leaking insulin at all the connection points and on the skin. Change the infusion set, the infusion site, the cartridge and the adapter. There is an occlusion (blockage) in the system.
  • Page 179: Appendix A: Technical Data

    Appendix A: Technical Data General Technical Data Maximum dimensions Approx. 82.5 × 56 × 21 mm (without adapter) (3.2 × 2.2 × 0.8 inches) Weight an empty insulin pump: approx. 80 g (2.8 ounces) insulin pump inclusive battery, full plastic cartridge and infusion set: approx.
  • Page 180 General Technical Data Barometric pressure During operation or storage in its shipping case: 50 to 106 kPA (500 to 1060 mbar) (do not exceed 3000 meters [10 000 feet] above sea level) During transport: 50 to 106 kPA (500 to 1060 mbar) Power supply One 1.5 volt AA Alkaline (LR6) or Lithium (FR6) battery or one 1.2 volt rechargeable NiMH AA (HR6) battery.
  • Page 181 General Technical Data Delivery Your Accu-Chek Spirit Combo insulin pump delivers insulin according to the programmed hourly basal rate, 24 hours a day. Minimal dose: 1/20 of the hourly basal rate, maximal 20 times per hour. Test interval: every 3 minutes.
  • Page 182 Flow (delivery rate) During infusion set priming and bolus: 0.2 U / sec. Maximum quantity delivered at ≤ 1.0 U a single fault condition Cartridge Accu-Chek Spirit 3.15 ml Cartridge System and Accu-Chek 3.15 ml Plastic Cartridge with a luer connection.
  • Page 183 General Technical Data Infusion sets Accu-Chek infusion sets with a luer-lock connector (see Appendix C). Accu-Chek Infusion sets are PVC and latex free. Data transfer Infrared interface Safety system Alert system, beeps, information on the display, vibrations, dual microprocessors. Your insulin pump is controlled by two microprocessors.
  • Page 184 Guidance and manufacturer’s declaration – electromagnetic emissions The Accu-Chek Spirit Combo insulin pump is intended for use in the electromagnetic environment specified below. The customer or the user of the Accu-Chek Spirit Combo insulin pump should ensure that it is used in such an environment.
  • Page 185 Guidance and manufacturer’s declaration – electromagnetic immunity The Accu-Chek Spirit Combo insulin pump is intended for use in the electromagnetic environment specified below. The customer or the user of Accu-Chek Spirit Combo insulin pump should assure that it is used in such an environment.
  • Page 186 Immunity test standard IEC 60601 test level * Compliance level Electromagnetic environment – guidance Voltage dips, short < 5 % UT (0.5 cycle) Not applicable interruptions and voltage 40 % UT (5 cycles) variations on power 70 % UT (25 cycles) supply lines <...
  • Page 187 Electromagnetic environment – guidance Portable and mobile RF communications equipment should be used no closer to any part of the Accu-Chek Spirit Combo insulin pump, including cables, than the recommended separation distance calculated from the equation applicable to the frequency of the transmitter.
  • Page 188 If the measured field strength in the location in which the Accu-Chek Spirit Combo insulin pump is used exceeds the applicable RF compliance level above, the Accu-Chek Spirit Combo insulin pump or the device, which contains it should be observed to verify normal operation.
  • Page 189 Accu-Chek Spirit Combo. The Accu-Chek Spirit Combo is intended for use in the electromagnetic environment in which radiated RF disturbances are controlled. The customer or the user of the Accu-Chek Spirit Combo can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications equipment (transmitters) and the Accu-Chek Spirit Combo as recom- mended below, according to the maximum output power of the communication equipment.
  • Page 190 Note At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies. Note These guidelines may not apply in all situations. Electro- magnetic propagation is affected by absorption and reflection from structures, objects and people. Note An additional factor of 10 / 3 is used in calculating the recommended separation distance to decrease the likelihood that mobile / portable communications equip- ment could cause interference if it is inadvertenly...
  • Page 191 The maximum deviation of the delivered amount (overall mean percentage flow error) for U100 insulin is ≤ ± 5 %*, where your Accu-Chek Spirit Combo insulin pump is located ≤ 1 m (3.3 feet) above / below the infusion site.
  • Page 192 For U100 insulin the maximum deviation of a maximum bolus is ≤ ± 5 % and the maximum deviation of a minimum bolus is ≤ ± 30 %*. * The measurements were made according to IEC 60601-2-24:1998 with a Accu-Chek Tender infusion set PC 17 / 110 at room temperature.
  • Page 193: Appendix B: Configuration Parameters

    Appendix B: Configuration parameters Maximum parameter Typical range available on range adjustable with the the pump and standard appropriate Accu-Chek parameter settings* PC Software Basal rate Hourly basal rate + max. TBR 62.5 U / h max. amount: 0 – 125.0 U / h (hourly basal rate combined with temporary basal rate, max.
  • Page 194 Maximum parameter Typical range available on range adjustable with the the pump and standard appropriate Accu-Chek parameter settings* PC Software Bolus Bolus amount per bolus delivery 25.0 U max. amount: 0 – 25.0 U bolus amount increments for Quick Bolus 0.5 U...
  • Page 195 Maximum parameter Typical range available on range adjustable with the the pump and standard appropriate Accu-Chek parameter settings* PC Software Battery type AA (LR6) ALKALINE, AA (FR6) LITHIUM or AA (HR6) NiMH rechargeable battery Key Lock turned on or off Automatic off turned on (1 –...
  • Page 196: Appendix C: Sterile Products And Accessories

    Comments Name Comments Accu-Chek Spirit 3.15 ml Plastic cartridges are single-use Accu-Chek Tender All Accu-Chek infusion sets are Cartridge System items only. Do not re-use cart ridges. Accu-Chek Ultraflex available in various tubing and Accu-Chek 3.15 ml Plastic Refer to the instructions for use Accu-Chek Rapid-D cannula lengths.
  • Page 197 Comments Name Comments Adapter Adapter and Battery Cover are part Carrying Systems All Accu-Chek Spirit Combo and Battery Cover of the Accu-Chek Spirit Combo Accu-Chek Spirit Carrying systems Service Pack and Accu-Chek are compatible with the Accu-Chek Spirit Combo Service Pack small.
  • Page 198: Appendix D: Abbreviations

    Appendix D: Abbreviations Abbreviation Meaning Abbreviation Meaning NiMH Nickel Metal Hydrid (used with rechargeable approx. Approximately batteries) Basal Rate rel. Relative Basal rate profile sec. Seconds Hour (s) Temporary Basal Rate International Electrotechnical Commission See I.U. incl. Inclusive or including U / h Amount of International Units of insulin Infrared...
  • Page 199: Appendix E: Beeps And Melodies

    Appendix E: Beeps and melodies Acoustic Signals All of the functions of the Accu-Chek Spirit Combo insulin pump Sound Function are supported by acoustic signals. The acoustic signals are beeps. DOWN KEY Frequency: 2486 Hz ±5 % Duration: Short sound...
  • Page 200 Sound Function Sound Function BASE STATE RUN (KEYS UNLOCKED) CARTRIDGE CONTENT <= 20 U Frequency: 2789 Hz ±5 % Frequency: f1 = 2090 Hz ±5 %, Duration: Short sound, short pause, f2 = 2486 Hz ±5 %, short sound, short pause, f3 = 2789 Hz ±5 % short sound Duration:...
  • Page 201 Sound Function Sound Function 50 U < CARTRIDGE CONTENT <= 100 U COPY BASAL RATE Frequency: f1 = 2090 Hz ±5 %, Frequency: 2486 Hz ±5 %, 3131 Hz ±5 %, f2 = 2486 Hz ±5 %, 2486 Hz ±5 % f3 = 2789 Hz ±5 % Duration: 3 x short sound...
  • Page 202 Sound Function Sound Function Reminder (REPEATING SEQUENCE) ERROR (REPEATING SEQUENCE) Frequency: f1 = 1043 Hz ±5 %, Frequency: f1 = 1043 Hz ±5 %, f2 = 1850 Hz ±5 % f2 = 1850 Hz ±5 % Duration: f1 medium sound, short Duration: f1 medium sound, short pause, f2 medium sound,...
  • Page 203 Sound Function Sound Function CONFIRMATION KEY PRESSED INITIAL POINT (SUBMENU, VALUE) Frequency: 2789 Hz ±5 % Frequency: 2820 Hz ±5 % Duration: Short sound Duration: Short sound, short pause, short sound CHANGED TO BASE STATE RUN BY OK KEY (KEYS UNLOCKED) SUPERVISOR SEQUENCE (GENERATED BY Sequence: A12, long pause, A14 SUPERVISOR)
  • Page 204 Tactile Signals Dedicated functions of the Accu-Chek Spirit Combo insulin pump are supported by tactile signals additionally. The tactile signals are vibrations which are generated by a vibra motor. Signal Function Signal Function WARNING AND Reminder SHORT VIBRATION (REPEATING SEQUENCE) Duration short vibration = 109.375 ms ±5 %...
  • Page 205: Display Symbols

    Appendix F: Symbols Display Symbols Icon Meaning Icon Meaning Basal Lock OFF Bluetooth ® Basal Lock ON Bluetooth Settings ® Bluetooth ® in Remote control Basal Rate mode Basal Rate Total Beep Test Battery Empty Cartridge Empty Battery Low Cartridge Full Beep Volume Cartridge Low Bluetooth...
  • Page 206 Icon Meaning Icon Meaning Clock Key Lock ON Data Transfer menu Pump timer (Call for update) Data Transfer (to start screen) Menu Settings Data Transfer running Multiwave Bolus Date Multiwave Bolus Immediate Date Format EU My Data Date Format US Pairing Failed Display contrast Pairing Successful...
  • Page 207 Icon Meaning Icon Meaning Pump Settings Stop Reminder Total Reminder Everyday Temporary Basal Rate Reminder OFF Therapy Settings Reminder Settings Time and Date Settings Reminder Single Units Display Orientation Units per hour Beep Vibrations Beep and Vibration Error Standard Bolus Warning...
  • Page 208 General Symbols Icon Meaning Icon Meaning Caution. Please refer to safety- Federal law (USA) restricts this related notes in the instructions for device to sale by or on the order use accompanying this product. of a physician. Bluetooth wireless technology ®...
  • Page 210: Appendix G: Warranty

    Roche Insulin Delivery Systems Inc. warrants the ACCU-CHEK ® Spirit Combo insulin pump (“insulin pump”) against defects in materials and workmanship for a period of four (4) years from the date of purchase. Proof of purchase date may be required.
  • Page 211 This warranty is valid only if the insulin pump is used in accord- SPECIAL OR PUNITIVE DAMAGES ARISING FROM OR IN ANY WAY ance with the user instructions. This warranty will not apply: CONNECTED WITH THE PURCHASE OR USE OF THE INSULIN If damage results from changes or modifications made to the PUMP.
  • Page 212: Glossary

    Your basal rate is delivered by your Accu-Chek Spirit Combo insulin pump, using Bluetooth ® wireless Accu-Chek Spirit Combo insulin pump according to the curve of technology. your personal basal rate profile or profiles. Adapter Basal rate profile The adapter connects the cartridge and infusion set.
  • Page 213 Cartridge compartment The term “Bluetooth” refers to a wireless technology that allows The chamber in your Accu-Chek Spirit Combo insulin pump for digital devices to transfer data at high speed. This technology is the cartridge. common in many portable devices such as portable PCs, Personal Digital Assistants and mobile telephones.
  • Page 214 Infusion set IPX8 according to IEC 60529 Infusion sets connect your Accu-Chek Spirit Combo insulin pump Protection against the effects of temporary immersion in water. to your body. The insulin is delivered from the cartridge through the Ingress of water in quantities causing damage is not possible...
  • Page 215 When your Accu-Chek Spirit Combo insulin pump is in STOP, it does not deliver any insulin. The insulin delivery is only stopped if an Error occurs or if your Accu-Chek Spirit Combo insulin pump is put into a necessary STOP, (e.g. when changing the cartridge, adapter or infusion set or for data transfer).
  • Page 216 STOP screen The STOP screen is the starting point for all functions that require that your Accu-Chek Spirit Combo insulin pump does not deliver insulin. Temporary Basal Rate Temporary increase or decrease of your basal rate profile in percentages (from 0 – 250 %) to match changing insulin needs due to increased or decreased activity level, illness or stress.
  • Page 217: Index

    Accessories ........1, 19 Accu-Chek Aviva Combo ......7, 133 Adapter .
  • Page 218 Electromagnetic immunity ......183 Language Electromagnetic radiation ......62 selecting .
  • Page 219 Reminders Tactile signals ........200 list of .
  • Page 220 Warnings and Errors ....... 151 confirming ........152 handling .
  • Page 221: Quick Reference Table For Errors, Warnings And Reminders

    Arrange for immediate replace- necessary. (loan pumps only) ment of your Accu-Chek OCCLUSION Change the cartridge and Spirit Combo insulin pump. infusion set. See section 2.6 to Contact your healthcare 2.9. professional for alternative insulin therapy options. END OF OPERATION Arrange for immediate replace- (“Backup”...
  • Page 222 What you have to do Warning What you have to do CARTRIDGE LOW Change cartridge before it is TBR CANCELLED Put your Accu-Chek Spirit completely empty. Combo insulin pump into RUN if necessary. Ensure the cancella- BATTERY LOW Replace the battery as soon as tion was intended and program possible.
  • Page 223 Reminder What you have to do LOANTIME WARNING In the MY DATA menu, check for Reminder Put your Accu-Chek Spirit Combo insulin pump into RUN if (loan pumps only) the remaining insulin pump loan time under the menu PUMP necessary.
  • Page 224 Bluetooth registration ® Please note that in Italy and France Bluetooth ® wireless technology is for indoor use only. Note: This equipment has been tested and found to comply with the limits for a Class B digital device, pursuant to part 15 of the FCC Rules.
  • Page 226: My Notes

    My notes...