If this happens ...
it means:

O
The drum compartment cover is not
properly closed.
Close the drum compartment cover. It
must close with an audible CLICK.
O
The ambient temperature is low and
the batteries are almost empty.
Move to a place where the tempera-
ture is between +10 °C and +40 °C
and wait for the temperature of the
meter to adjust to this temperature.
Insert batteries (new or recharged).
O
Condensation has affected the elec-
tronics.
Allow the meter to dry slowly.
O
The meter is defective.
Please contact the Accu-Chek
Customer Care Line.
182
Symbols, error messages and troubleshooting
If this happens ...
it means:
The meter is turned on,
The display is defective.
but the display is blank.
Please call Accu-Chek Customer Care at
1-800-363-7949.
The clock has stopped or
The meter was exposed to a temperature
the clock is slow.
below –10 °C and the batteries are begin-
ning to freeze.
Turn the meter off. Move to a place where
the temperature is between +10 °C and
+40 °C and wait for the temperature of
the meter to adjust to this temperature.
O
When the meter is turned
The meter was exposed to a tempera-
on, the date is shown as
ture below –10 °C and the batteries
0-00 and the time is
are beginning to freeze.
shown as 0:00.
Press the Test button to turn the meter
off. Move to a place where the tem-
perature is between +10 °C and
+40 °C and wait for the temperature
of the meter to adjust to this tempera-
ture.