6.5

Announcement of warnings and error messages

When the batteries will soon be exhausted or the last test strip in a
drum has been used, the meter issues an acoustic warning. This con-
sists of two short beeps sounded three times in succession
- - - - - -
(
).
The warning is given at different times.
O
Batteries are almost exhausted (the battery symbol is displayed,
see Chapter 9):
This acoustic warning is sounded when you turn the meter on.
O
You have used the last test strip in the drum (
flash, see Chapters 2.3 and 12.2):
The acoustic warning is given after you have pressed the Test button
to turn the meter off.
If error messages (
with a number or the
E
displayed, these are announced by two short beeps sounded four times
- - - - - - - -
in succession (
once and is not repeated. The same sequence of beeps is sounded for
all error messages, i.e. they are not differentiated acoustically.
128
and drum symbol
End
, see Chapter 12.2) are
EEE
). This sequence is sounded only
Acoustic Mode
7

Checking your meter

To ensure your results are always dependable, you should periodically
check the performance of your meter.
Do this by performing a performance check using the Accu-Chek Com-
pact Autocontrol control solutions, each time
O
after you have opened a new pack of test strip drums,
O
after you have replaced the batteries.
O
after your meter has displayed the bottle symbol
perform a performance check,
O
after you have cleaned the measuring window of the meter,
O
if your result does not agree with how you feel.
Please call Accu-Chek Customer Care at 1-800-363-7949 to obtain the
control solutions.