Display messages and troubleshooting guide
Display
82
82
11:23 Uhr
Seite 82
Cause
• A test error has occurred.
• There is electromagnetic
interference.
• You forcibly removed the
test strip during testing.
• You bent the strip while
applying blood or during
measurement.
• The optical system cover
may not be closed properly.
The test strip is stuck in the
guide.
Action
• Repeat the test.
• Move away from possible sources of interference
(see Chapter 5.3 "Testing conditions").
• Repeat the test with a new strip, this time ejecting it by
means of the ON/OFF button.
• Repeat the test with a new strip.
• Ensure that the optical system cover is firmly in place.
Pull the strip out of the monitor.
Carefully open the drum compartment and remove the drum
without turning it.
Reinsert the drum.
Close the drum compartment.
Keep the "use-by" period in mind when reusing the drum.