Display messages and troubleshooting guide
Display
or
74
74
11:23 Uhr
Seite 74
Cause
The bar code information
stored from the test strip
drum has been lost.
There is no test strip drum in
the monitor.
You inserted a test strip drum
that has already been used.
Action
Open the drum compartment and close it again.
The display showing the 90-day "use-by"
period will be lost (see Chapter 3).
Insert a test strip drum.
The monitor cannot determine how long the test strip drum
has been outside the monitor, or whether it is still usable. If
you are certain that the 3-month "use-by" period has not been
exceeded, you may continue to use the drum. Perform a quality
control check. The result of the quality control check will be
stored in memory along with "CTRL", and the drum symbol will
remain displayed until a new drum is inserted. If you do not
perform a quality control check, the flashing drum symbol and
the flashing "CTRL" will appear with every new test until a new
drum is inserted. Until this point, all test results will be stored
along with the flashing drum symbol and "CTRL".
Caution
75