Display messages and troubleshooting guide
Display
mmol/L
A dot appears in the control
window at the back of the
monitor.
72
72
11:23 Uhr
Seite 72
Cause
Temperature was outside the
indicated range during test-
ing.
It is too warm or too cold for
the test, i.e. warmer than plus
45°C or colder than plus 5°C.
There was not enough blood
on the strip.
The test strip drum in the
monitor is empty.
Action
The displayed test result may only be used for reference
purposes. It is saved along with the thermometer symbol.
To perform an accurate test, let the monitor and test strips
slowly cool down or warm up, and repeat the test. The
optimum measuring temperature is between plus 10°C and
plus 40°C.
No test result is displayed or saved. Let the monitor and test
strips slowly cool down or warm up, and repeat the test. The
optimum measuring temperature is between plus 10°C and
plus 40°C.
Repeat the test with a larger drop of blood.
Insert a new drum. If the message is still displayed, please
contact our Diabetes Care Info-Line (for details see Chapter 12
"Customer support and service").