11
Display messages
Display
Drum symbol and "E-2"
flash when the second
or third drum of a
package is inserted.
During subsequent tests
the drum symbol flashes.
84
Cause
The bar code on the drum
is damaged.
Action
Press the SET button. The meter uses the bar
code from the first drum. The error message dis-
appears. "CTRL" flashes in the display, requesting
you to perform a quality control check. You can
use all 17 test strips. The drum symbol continues
to flash in the display until a new drum is inserted.
All test strip drums in the same package have
a bar code unique to that package. Do not try
to code the meter using drums from a
different package. Check you are using the
correct drums by comparing their batch
(or lot) codes. The batch code is preceded by
the symbol
LOT
. If the message is still
displayed, please contact our Accu-Chek
Blood Glucose Meter Careline (see Chapter
12 "Customer support and service").