11
Display messages
Display
80
Cause
The bar code information
stored from the test strip
drum has been lost.
There is no test strip drum in
the meter.
You inserted a test strip drum
that has already been used.
Action
Open the drum compartment and close it again.
Warning! The display showing the 90-day
"use-by" period will be lost (see Chapter 3).
Insert a test strip drum.
The meter cannot determine how long the test
strip drum has been outside the meter, or
whether it is still usable. If you are certain that the
3-month "use-by" period has not been exceeded,
you may continue to use the drum. Perform a
quality control check. The result of the quality
control check will be stored in memory along with
"CTRL", and the drum symbol will remain
displayed until a new drum is inserted. If you do
not perform a quality control check, the flashing
drum symbol and the flashing "CTRL" will appear
with every new test until a new drum is inserted.
Until this point, all test results will be stored along
with the flashing drum symbol and "CTRL".
Warning!
81