Accu-Chek Compact Reference Manual page 41

Blood glucose system
Hide thumbs Also See for Compact:
Table of Contents

Advertisement

11
Display messages
Display
mmol/L
A dot appears in the
control window at the
back of the meter.
78
Cause
Temperature was outside
the indicated range
during testing.
It is too warm or too cold for
the test, i.e. warmer than plus
45°C or colder than plus 5°C.
There was not enough blood
on the strip.
The test strip drum in the
meter is empty.
Action
The displayed test result may only be used for
reference purposes. It is saved along with the
thermometer symbol. To perform an accurate test,
let the meter and test strips slowly cool down
or warm up, and repeat the test. The optimum
measuring temperature is between plus 10°C
and plus 40°C.
No test result is displayed or saved. Let the
meter and test strips slowly cool down or warm up,
and repeat the test. The optimum measuring tem-
perature is between plus 10°C and plus 40°C.
Repeat the test with a larger drop of blood.
Insert a new drum.
If the message is still displayed, please call our
Accu-Chek Customer Careline
(see Chapter 12 "Customer support and service").
79

Advertisement

Table of Contents
loading

Table of Contents