Power supply
2 batteries (type AAA, LR 03, AM 4 or
Micro) or 2 NiMH rechargeable batteries
(type AAA)
Battery life
approx. 1000 results (fewer in Acoustic
Mode, owing to higher battery load, or
when rechargeable batteries are used)
Interface
Infrared window
Safety class
III
LED/IRED
Class 1
Electromagnetic
This device meets the electromagnetic
compatibility
immunity requirements according to
EN ISO 15197 Annex A. The chosen
basis for the immunity tests (on elec-
trostatic discharge) was basic standard
IEC 61000-4-2.
In addition it meets the electromagnetic
emissions requirements according to
EN 61326. Its electromagnetic emis-
sion is thus low. Interference from other
electrically driven equipment is not to
be anticipated.
164
Performance evaluation
The Accu-Chek Compact Plus system complies with the requirements of
EN ISO 15197.
Technical data
Technical data
The performance data for the
Accu-Chek Compact Plus system
(Accu-Chek Compact Plus meter to-
gether with Accu-Chek Compact test
strips) were determined using capillary
blood from diabetic patients (method
comparison, accuracy), venous blood
(repeatability) and control solutions (re-
producibility). The method comparison
was conducted by comparing the re-
sults from the system with those of the
hexokinase method with deproteination
using an automatic analyzer. The hexo-
kinase method is traceable to an NIST
standard (traceability).