This message/symbol
and means:
appears...

The bar code on the test strip drum
is damaged.
If you inserted the first drum from
a pack, remove the drum from the
meter and return it to its container.
Insert the second drum from the
same pack in the meter and use
all 17 test strips. When it is time to
replace the drum, insert the first
drum with the damaged bar code.
When the error message is displayed
again, press the S button. The meter
uses the bar code from the preced-
ing drum. The drum and bottle sym-
bols flash in the display. The meter
requests you to perform a quality
control check. You may use all 17
test strips for testing. The drum sym-
bol continues to flash in the display
until you have inserted a new drum.
148

Symbols, display messages and troubleshooting
This message/symbol
and means:
appears...

Symbols, display messages and troubleshooting
If you inserted the second or third
drum from a pack, press the S but-
ton. The meter uses the bar code
from the preceding drum. The drum
and bottle symbols flash in the
display. The meter requests you to
perform a quality control check. You
may use all 17 test strips for testing.
The drum symbol continues to flash
in the display until you have inserted
a new drum.
If you inserted a test strip drum from
single-drum pack, you can no longer
use this drum. Insert a new drum.