Accu-Chek Compact Plus Reference Manual page 75

Blood glucose monitoring system
Hide thumbs Also See for Compact Plus:
Table of Contents

Advertisement

This message/symbol
and means:
appears...
Although you performed a
An assignment error has occurred. The
quality control check, the
result is correct, but has been saved as
bottle symbol is not dis-
a blood glucose result.
played.
Press the S button twice so that the
bottle symbol is displayed.
If the meter has already been
switched off, you can no longer
make the correction. You cannot
change results you have retrieved
from memory with the M button.
146
Symbols, display messages and troubleshooting
This message/symbol
and means:
appears...
Symbols, display messages and troubleshooting
The bar code information from the
test strip drum has been lost.
Press the Test button to turn the
meter off. Open and close the drum
compartment.
The meter then rotates the drum
and reads the bar code. The drum
symbol and the bottle symbol flash
in the display when the next test is
performed. Perform a quality control
check. You may use up all the test
strips remaining in the drum. The
90-day use-by period will no longer
be properly calculated for this drum.
The drum symbol continues to flash
in the display until you have inserted
a new drum.
147
10
