This message/symbol
and means:
appears...
(cont.)
The infrared windows are too far
apart or not properly facing one an-
other.
Reduce the distance between the
two infrared windows to 10-20 cm
and position them so they are fac-
ing one another. Start the download
again.
A software error occurred on the PC,
handheld computer or the
Accu-Chek Smart Printer System.
Read the manual for the software
or the Accu-Chek Smart Printer
System.
144
Symbols, display messages and troubleshooting
10

Technical data

Meter type
Accu-Chek Compact Plus
Catalogue No./Serial No.
See type plate on back of meter
Test principle
Determination of glucose in fresh capil-
lary blood by reflectance photometry.
When using different specimen material
please refer to the package insert that
came with the Accu-Chek Compact test
strips.
Measuring range
0.6–33.3 mmol/L (10–600 mg/dL)
Sample size
approx. 1.5 µL (1 microlitre =
1 thousandths of a millilitre)
Measuring time
approx. 5 seconds (depending on
concentration)
System operating
temperature
+10 °C – +40 °C
Storage temperature
Meter
without batteries and test strip drum:
-25 °C – +70 °C
with batteries but without test strip
drum: -10 °C – +50 °C
with batteries and test strip drum:
+2 °C – +30 °C
Lancing device
-25 °C – +70 °C
Technical data