CALIBRATION PATH

The calibration is performed in Kg or Lb, according to the units of measure used upon entering
into the calibration path.
Operator Action
1.
Press ON/OFF button and immediately press
and hold HOLD/RELEASE button for 3-4
seconds
2.
Press ENTER
3.
Using
keys and/or the keypad, set the
calibration load to at least 200 lbs (for best
results and greater accuracy, use a 400 lb
weight). Press ENTER
4.
Please clear the stretcher and press ENTER
5.
Please wait without touching or shaking the
scale until the zero calibration process is
finished
6.
Load the stretcher with the required weight and
press ENTER
7.
Please wait without touching or shaking the
scale until the calibration process is finished
8.
Press ENTER
9.
Please remove weight from the stretcher
10. Please wait until the scale resumes normal
operation
37

EXPLODED VIEW OF SCALE

38
39
40
Calibration
Process
Enter load weight
Zero calibration
Please clear the scale
Zero calibration
Please wait
Weight calibration
Weight calibration
Please wait
Calibration
factor: X.XXXXX
Calibration
Please clear the scale
Rebooting
Please wait
Health O Meter
Pro Plus
42
41
43
44
52
25
Display
400.0
Put: XXX.X
45
46
51
50
47
48
49

Advertisement

Table of Contents
loading

This manual is also suitable for:

Pro plus 2000kl

Table of Contents