CALIBRATION PATH

The calibration is performed using kg or lb, according to the units of measure used upon
entering into the calibration path.
1.
Press ON/OFF button and immediately press and
hold HOLD/RELEASE button for 3-4 seconds
2.
Press ENTER
3.
Using
keys and/or the keypad, set the calibration
load to at least 200 lbs (for best results and greater
accuracy, use a 600 lb weight). Press ENTER
4.
Please clear the scale platform and press ENTER
5.
Please wait without touching or shaking the scale
until the zero calibration process is finished
6.
Load the scale platform with the required weight and
press ENTER
7.
Please wait without touching or shaking the scale
until the calibration process is finished
8.
Press ENTER
9.
Please remove weight from the scale platform
10. Please wait until the scale resumes normal operation
Operator Action

EXPLODED VIEW OF SCALE

Display Assembly
Calibration
Process
Enter load weight
600.0
Zero calibration
Please clear the scale
Zero calibration
Please wait
Weight calibration
Put: XXX.X
Weight calibration
Please wait
Calibration
factor: X.XXXXX
Calibration
Please clear the scale
Rebooting
Please wait