13

Technical data

Meter type
Accu-Chek Compact Plus (Model GT)
Catalogue No./Serial No. See type plate on the back of the meter
Test principle
Determination of glucose in fresh capillary
blood by reflectance photometry. When using
different specimen material please refer to
the package insert that came with the
Accu-Chek Compact test strips.
Blood glucose concentrations may be mea-
sured in whole blood or in plasma. Although
you always draw up whole blood with the test
strip, your meter displays blood glucose
results that relate either to whole blood or to
plasma. To see whether your blood glucose
meter displays results relating to whole blood
or to plasma, please read the package insert
that came with your test strips. There you will
also find information on how the system
works, on the test principle and on reference
methods.
Measuring range
0.6–33.3 mmol/L (10–600 mg/dL)
Sample size
approx. 1.5 μL (1 μL (microlitre) =
1 thousandth of a millilitre)
196
Measuring time
System operating
temperature
Storage temperature
Meter
Lancing device
Humidity
during testing
during storage
Altitude
Memory
Dimensions
without lancing device
with lancing device
Technical data
Technical data
approx. 5 seconds (depending on concentra-
tion)
+10 °C to +40 °C
without batteries and
without a test strip drum: -25 °C to +70 °C
with batteries and
without a test strip drum: -10 °C to +50 °C
with batteries and
with a test strip drum: +2 °C to +30 °C
-25 °C to +70 °C
20 % to 85 % relative humidity
20 % to 85 % relative humidity
Sea level to 4000 m
500 results together with time and date, aver-
ages, highest and lowest value for 7, 14 and
30 days
121 × 57 × 32 mm
125 × 64 × 32 mm
13