Accu-Chek Compact Plus User Manual page 94

Blood glucose monitoring system
Hide thumbs Also See for Compact Plus:
Table of Contents

Advertisement

Error message
Meaning
The meter rotates the drum and reads
the bar code. When you next turn on
the meter to perform a test, the drum
symbol and flashing bottle symbol are
displayed. Perform a performance
check. You may use all the test strips
remaining in the drum. The 90-day
use-by period will no longer be calcu-
lated for this drum. The drum symbol
will therefore be displayed each time
you perform a test until you insert a
new test strip drum.
The bar code on the test strip drum is
damaged.
You can no longer perform tests with
test strips from this drum. Insert a
new drum.
186
Symbols, error messages and troubleshooting
Error message
Meaning
The measuring window is dirty.
Press the Test button to turn the meter
off. Clean the measuring window (see
Chapter 8.1) and repeat the test with a
new test strip.
A test strip is stuck in the meter.
Hold the meter upright so the test strip
is pointing downwards. Press the Test
button to turn the meter off and allow
the test strip to drop out. If the test
strip does not drop out, remove it
manually, even if you feel some resis-
tance.
The test strip drew up too little blood
or control solution.
Turn the meter off and then on again
by pressing the Test button. Repeat
the test with a new test strip and a
larger amount of blood or control solu-
tion.
Symbols, error messages and troubleshooting
12
187

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents