Accu-Chek Compact Plus User Manual page 92

Blood glucose monitoring system
Hide thumbs Also See for Compact Plus:
Table of Contents

Advertisement

If this happens ...
it means:
The drum compartment cover is not
properly closed.
Close the drum compartment cover. It
must close with an audible CLICK.
The ambient temperature is low and
the batteries are almost empty.
Move to a place where the tempera-
ture is between +10 °C and +40 °C
and wait for the temperature of the
meter to adjust to this temperature.
Insert batteries (new or recharged).
Condensation has affected the elec-
tronics.
Allow the meter to dry slowly.
The meter is defective.
Please contact the Accu-Chek
Customer Care Line.
182
Symbols, error messages and troubleshooting
If this happens ...
it means:
The meter is turned on,
The display is defective.
but the display is blank.
Please call Accu-Chek Customer Care at
1-800-363-7949.
The clock has stopped or
The meter was exposed to a temperature
the clock is slow.
below –10 °C and the batteries are begin-
ning to freeze.
Turn the meter off. Move to a place where
the temperature is between +10 °C and
+40 °C and wait for the temperature of
the meter to adjust to this temperature.
When the meter is turned
The meter was exposed to a tempera-
on, the date is shown as
ture below –10 °C and the batteries
0-00 and the time is
are beginning to freeze.
shown as 0:00.
Press the Test button to turn the meter
off. Move to a place where the tem-
perature is between +10 °C and
+40 °C and wait for the temperature
of the meter to adjust to this tempera-
ture.
Symbols, error messages and troubleshooting