Possible Causes Of Error During The Performance Of Performance Checks - Accu-Chek Compact Plus User Manual

Blood glucose monitoring system
Hide thumbs Also See for Compact Plus:
Table of Contents

Advertisement

Subsequently hold the meter directly above a waste bin with the test
strip pointing downwards.
Press the Test button to turn the meter off.
The used test strip drops out.
Before the meter turns off, it shows you the number of test strips
remaining in the drum (see Chapter 2.3.2).
Wipe the tip of the dropper with a clean, dry paper towel and close
the bottle tightly.
Notes
The control solution you apply to a washable object must come
i
from a suspended drop. Do not use the drop if it has run down
the outside of the tip of the dropper. This may lead to results that
are too high.
Control solution that is to be drawn up by the test strip must just
i
have been freshly applied to the object. The control solution must
not be left exposed on the surface for any length of time, or it
may give results that are too high.
Results flagged as performance checks with the bottle symbol
i
are not included in the calculation of averages and lowest and
highest results.
136
Please refer also to the notes in Chapters 4.3 and 4.4.
7.3
If the result is outside the stated concentration range, perform a second
performance check. If the result of this second check is outside the con-
centration range as well, please check the following points:
Checking your meter
Checking your meter
You can dispose of empty or only partly used bottles of control
i
solution together with your regular household waste.
Possible causes of error during the performance of perfor-
mance checks
Did you perform the performance check as described above?
Did you wipe the tip of the dropper before you applied control
solution to the surface of the washable object?
Did you use a suspended drop of control solution?
Was there prolonged contact between the control solution and
the washable object?
Did you bend the test strip either before or while you were
drawing up control solution?
Did you wait for the test strip symbol and drop symbol to flash
in the display before drawing up control solution with the test
strip?
7
yes
yes
yes
no
no
yes
137

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents