Set Up / Batteries - Health O Meter PROPLUS 4024 User Instructions

Portable, adjustable patient wheelchair scale

Advertisement

Available languages

Available languages

SET UP / BATTERIES

The ProPlus® Portable, Adjustable Wheelchair Scale Model 4024 is shipped with the Menu Lock
function enabled. The Menu Lock function disables certain buttons on the scale's keypad, thus
preventing unauthorized users from making changes to your scale settings. The Menu Lock can be
bypassed temporarily to allow you to customize your scale with the functionality you prefer. Please
refer to the ProPlus® Display Module User Manual for instructions on how to bypass the Menu Lock.
Preparing the Scale for Use
1.
Remove protective plastic film from keypad and display.
Place batteries in the battery holder (see ―Batteries‖ information, below).
2.
Plug the scale's AC adapter into the power source.
3.
Press the ON/OFF button to turn the scale on. The display will show ―Health o meter
4.
ProPlus‖ and then ‖0.00 lb / 0.0 oz‖. Note: To adjust the display backlight and/or contrast, please refer
to the ProPlus® Display Module User Manual.
5.
Place a wheelchair on the scale to register a 0 weight on the display. Press the TARE key, if
necessary—see page 4 of this manual for more information regarding TARE.
Place a weight that is WITHIN the scale's capacity on the wheelchair (see page 3 of this manual for
6.
more information). The display should read ―WEIGHING‖ until the weight calculation is complete and
the weight is displayed.
7.
Press the KG/LB button to select the weight mode (pounds or kilos). To change the display mode in
pounds to fractions or decimals, please refer to the ProPlus® Display Module Manual.
8.
Press the REWEIGH button; the scale should perform the weighing process again.
Remove the weight from the wheelchair. The scale will return to zero and the display will read ―ZERO‖
9.
on the left upper side of the screen along with ‖0.00 lb / 0. 0oz‖.
Disconnect the scale's AC adapter from the power source--the scale will switch to battery power if
10.
there are batteries in the scale.
Note: If the set up procedure failed, refer to the troubleshooting instructions on page 10. If the
problem is still not corrected, contact qualified service personnel.
Replacing / Inserting Batteries
1.
Unplug the scale.
2.
Remove the battery cover from the display assembly—refer to Figure 4 (A) above.
3.
Disconnect battery holder cable connector from the scale-battery connector (B).
4.
Carefully remove the battery holder (C) by sliding it out of the display assembly.
5.
Replace the batteries with new ones. We recommend EVEREADY Energizer® e2™ batteries.
6.
Carefully slide the battery holder (C) into the display assembly.
7.
Plug the battery holder cable connector to the scale-battery connector (B).
8.
Attach the battery cover to the display assembly and install the screw (A).
Warning: If the scale will not be used for some time, remove batteries to avoid a safety hazard. Disposal
of batteries should be done according to the national, regional or local regulations which apply to you for
disposal of batteries.
Figure 4 - Batteries