Download Print this page

Maintenance / Calibration - Health O Meter PROPLUS 4024 User Instructions

Portable, adjustable patient wheelchair scale

Advertisement

Available languages

Available languages

QUICK START INSTRUCTIONS (CONTINUED)
Body Mass Index is another name for Quetelet Index, a formula that is a convenient and reliable indicator
of obesity. NOTE: The scale will not calculate BMI for a patient that is 26.4 pounds (12 kg) or less. If a
height is not entered for the patient within 30 seconds of pressing the BMI button, the scale will return to the
normal weighing mode.
1. Complete steps 1 to 4 above for ―Weighing A Patient‖.
2. After obtaining the patient's weight on the scale, press the BMI (2) button while the scale is on.
3. The scale will prompt you to enter the patient's height. Use the keypad to enter the height.
NOTE: When weighing in pounds, the height is entered in ¼‖ increments. To increase the fractional
height of the patient, press ―1‖ for ¼‖, ―2‖ for ½‖ and ―3‖ for ¾‖. When weighing in kilograms, the height
must be entered in 0.1 centimeter increments.
4. Once height is keyed in, press ENTER.
5. The display will show the patient's BMI (Body Mass Index).
6. Press EXIT to return to the normal weighing mode of the scale.
MAINTENANCE
GENERAL
This section provides instructions for maintenance, cleaning, troubleshooting and operator
replaceable parts for the ProPlus
described in this section should be performed by qualified service personnel.
MAINTENANCE
Before first use and after periods of non-use, check the scale for proper operation and function. If
the scale does not operate correctly, refer to qualified service personnel.
1.
Check overall appearance of the total scale for any obvious damage, wear and tear.
2.
Inspect AC adapter for cord cracking or fraying or for broken or bent prongs.
CALIBRATION
This scale has been factory calibrated and does not require calibration prior to use. For
maintenance, the scale can be calibrated--please refer to your Display Module User Manual or
call Health o meter® Professional Technical Support at (1) 800-815-6615 or (1) 708-377-0600.
CLEANING
Proper care and cleaning is essential to ensure a long life of accurate and effective operation.
Disconnect the scale from the AC power source.
1.
Clean all external surfaces with a clean damp cloth or tissue. Mild antimicrobial or antibacterial soap
and water solution may be used. Dry with a clean soft cloth.
2.
Do not immerse the scale into cleaning or other liquid solution. Do not gas or steam sterilize the scale,
nor place any part of it through an autoclave.
3.
Do not use Isopropyl Alcohol or other solutions to clean the display surface.
4.
Do not use abrasive cleaners
CALCULATING BMI
®
Scale Model 4024. Maintenance operations other than those
.
10

Advertisement

loading