This scale has been factory calibrated, and does not require calibration prior to use.
If required, the scale can be calibrated. Calibration of your scale is performed using kilograms
(KG) or pounds (LB), according to the units of measure used upon entering into the calibration
path. When calibrating in pounds, a total of 300 lbs is required (models with date codes listed
below require 500 lb), applied in increments of 100 lb and 200 lb. If calibrating in kilograms, a
total of 120 kg is required (or 200 kg for models with date codes listed below), applied in
increments of 40 kg, and 80 kg. THE 500KG CAN ONLY BE CALIBRATED IN KILOGRAMS.
Note: Only weights certified and traceable to national standards are to be used for
calibration procedures.
*For scales with these date codes 500 lb (200kg) of weights are required for calibration.
2215, 2315, 2715, 2815, 3015, 3115, 3215, 3415, 3515, 3715, 3815, 3915, 4115, 4215, 4315,
4515, 4615, 4715, 4915, 5015, 5215, 0316, 0916, 1016, 1116, 1216, 1316, 1516, 1616, 1816
With the scale off, press and hold the LB/KG button (500KL) or the blank
blue button (500KG) and ZERO buttons at the same time, then press the
ON/OFF button to turn on the scale. Release all three buttons when "CAL"
appears on the display. Numbers will appear indicating that you can begin
calibration.
When calibration numbers stabilize, press the ZERO button. "C-100" (for
LB mode) or "C-40" (for KG mode) will flash on the display, followed by "0"
or "1".
Place 100 lbs or 40 kg on the scale. Numbers will appear on the display.
When calibration numbers stabilize, press the ZERO button. "C-200" (for
LB mode) or "C-80" (for KG mode) will appear on the display, followed by
"0" or "1".
Place an additional 100 lbs or 40 kg on the scale, for a total of 200 lbs or
80 kg. Numbers will appear on the display. When calibration numbers
stabilize, press the ZERO button. "C-300" (for LB mode) or "C-120" (for KG
mode) will appear on the display, followed by "0" or "1".
Place an additional 100 lbs or 40 kg on the scale, for a total of 300 lbs or
120 kg. Numbers will appear on the display. When calibration numbers
stabilize, press the ZERO button. Calibration procedure is now complete
and "END" will appear on the display. For models with date codes listed
above the display will show "C-500" (for LB mode) or "C-200" (for KG
mode) will appear on the display, followed by "0" or "1".
Place an additional 200 lbs or 80 kg on the scale, for a total of 500 lbs or
200 kg. Numbers will appear on the displaly. When calibration numbers
stabilize, press the ZERO button. "END" will appear on the display,
followed by "OFF". The scale will automatically shut off.
P/N UM500KL 500KG Rev 20190408_Web
ACTION
8
DISPLAY SHOWS
"CAL" followed by
numbers
Numbers followed
by "C-100" or
"C-40" then "0" or
"1"
Numbers followed
by "C-200" or
"C-80" then "0" or
"1"
Numbers followed
by "C-300" or
"C-120" then "0" or
"1"
"END"
umbers
or n
followed by "C-500"
or "C-200" then "0"
or "1"
"XXXX" followed
by "END" then
"OFF"