Setup Instructions
3.5 Connection of Auxiliary Devices
Content in this section
Note
Always observe the instructions given in the manual of the peripheral/auxiliary device.
Warning
After each installation, the leakage currents have to be measured according to IEC 60601-1, IEC 62353 or
other relevant standard.
Caution
Please observe that some printers may not be medical devices! If Bluetooth-, Line- or Laser Printers are no
medical devices, they have to be located outside of the typical patient environment. Examples for typical
patient environments can be found in standard IEC 60601 (see illustrations below).
For more detailed Safety Considerations when connecting auxiliary devices to the Voluson E-Series system,
Note
please review:
3-14
Section 1.5.5 "Auxiliary Devices Safety" on page 1-21
.
Voluson E-Series Service Manual
5539550APB Revision 6