Replacement Of The Cable Holder; Cable Holder - Replacement Procedure; Replacement Of The Probe Holder (Kit); Probe Holder (Kit) - Replacement Procedure - GE Voluson E Series Service Manual

Hide thumbs Also See for Voluson E Series:
Table of Contents

Advertisement

Replacement Procedures

8.9 Replacement of the Cable Holder

Manpower
One person, 3 minutes
Tools
Phillips screwdriver PH1 and PH2

8.9.1 Cable Holder - Replacement Procedure

1.
Loosen 2 screws below the control console and then remove the Cable Holder.
Figure 8-26 loosen 2 screws and remove cable
2.
Place the new Cable Holder at its original position and fasten it with 2 screws.

8.10 Replacement of the Probe Holder (Kit)

Manpower
One person, 1 minute
Tools
none

8.10.1 Probe Holder (Kit) - Replacement Procedure

1.
Simply pinch the clip and then pull out the Probe Holder to be replaced.
2.
Insert the new Probe Holder from the kit.
8-22
holder
Figure 8-27 remove the probe holder
Voluson E-Series Service Manual
5539550APB Revision 6