Replacement Procedures
8.7 Replacement or Activation of Options
For a list of available SW options and detailed description, refer to
Note
Options" on page 5-12
It might be possible that some probes, options or features are NOT available
•
in some countries.
•
at the time of release of this Service Manual.
8.7.1 How to activate Options by means of a "Demo Key" or a "Permanent Key"
1.
Press the Utilities key on the control console.
2.
In the "Utilities" menu touch the
3.
On the right side of the screen select
1 D = Demo (Option is activated for demo and expires at date shown in the "Valid" column)
2 Permanent Key
4.
Position the cursor inside the input field desired and press the right/left trackball key.
5.
If one exists, clear/edit the current Permanent Key.
6.
Enter the encrypted Permanent Key with the keyboard and click on
7.
Click the
After activating a Permanent Key, restart (turn off and on) the Voluson E-Series system.
Note
At EC310 the format of the Permanent key has been changed.
New Format from EC310 onwards:
63XQ-4DE4-Ek1D-A3CFT-JQ0C-YSFF
e.g.,
Characters are case sensitive !!
Hyphens (-) must not be entered. They are
self-created.
8-16
Setup button to invoke the setup desktop on the screen.
Figure 8-20 Setup - Administration - Options page (e.g., Voluson E10)
I = Inactive (Option is not activated)
P = Permanent (Option is permanently activated, i.e.purchased)
d = deactivated (Option is inactive)
Save&Exit button.
Section 5.1.4 "Description of Software
Administration and then click the Option tab.
Previous Format still valid until EC300:
02148EBCA0069003EF6F980F
e.g.,
Characters are not case sensitive.
Submit . (Code will be checked.)
Voluson E-Series Service Manual
5539550APB Revision 6