7.8.7 Network Troubleshooting

7.8.7.1 No Connection to the Network at all
Gray = Cable disconnected or no network signal on a connected cable.
Green = Cable connected to a network. Does not imply proper network settings.
1.
Check that the network cable between the Voluson E-Series system and the wall network is connected
and well seated in both ends. (Use a network cable that is known to be OK.)
2.
Connect a network cable between the system and a PC by either using a hub or a cross-over cable.
Try to ping from system to IP address on PC. If OK, hardware connection inside the system is OK.
7.8.7.2 GE remote service connection
Visual information about GE remote status is shown in the status area on the right bottom of the screen. At
Note
EC330 (BT18) systems the trackball functions are displayed in the trackball status area of the screen.
Gray = Idle State
Yellow = Access
pending
Red = Disrupted
State
Gray + envelope
Voluson E-Series Service Manual
5539550APB Revision 6
Remote access is inactive.
EC330 only: Move the cursor over the
Contact :
Section 7.2 "Request for Service (RFS)" on page 7-5
Configuration :
Section 3.13.8 "InSite Configuration" on page 3-70
Menu : Displays the Remote Access (Disruptive Mode) menu
Remote Access connection is active, but Disruptive Mode and Virtual Console
Observation (VCO) is not yet enabled.
Remote Access is active. All processes [UL_VNC and UL_CSD] are active. In this
state the Voluson E-Series system should NOT be used clinically.
New Software Update Package is available for download and installation.
Section 8.2.3 on page 8-10
Operation see
Diagnostics/Troubleshooting
GE icon to show trackball functions
.