Functional Checks
4.6 Peripheral Checks
Check that peripherals work as described below:
Step
1
Press the Freeze key.
2
Press the remote key (P1, P2, P3, P4, P5 or P6),
which is assigned to the BW printer.
3
Press the remote key (P1, P2, P3, P4, P5 or P6),
which is assigned to the color printer.
4
Press the remote key (P1, P2, P3, P4, P5 or P6),
which is assigned to recorder control.
Verify basic operations of further auxiliary devices.
4.6.1 ECG Check Out
Connect the ECG preamplifier and check:
Step
1
Press the Utilities key on the control console and then
touch the
4-24
Table 4-1 Peripheral Checks
Page + Task
Table 4-2 Peripheral Checks
Page + Task
ECG button to display the "ECG" menu.
Expected Result(s)
Stop image acquisition.
The image displayed on the screen is printed on the
Black & White printer.
The image displayed on the screen is printed on the
Color printer.
Recording starts/stops.
Expected Result(s)
It will display a curve along the bottom edge of the
image sector.
Voluson E-Series Service Manual
5539550APB Revision 6