The PP006A is a high impedance passive probe with a wide range of probing accessories. It
is designed to give high impedance probing capability to a circuit when attached to a LeCroy
Oscilloscope.
System BW (-3dB)
System Attenuation
System Input Resistance
Compensation Range
Cable Length
Max. non-destructive input voltage : 600 V (DC + Peak AC) (see derating curve)
Probe Sense
A ccessories Table
① Sprung hook
② Ground lead (11 cm)
③ Ground lead on probe tip
④ IC Insulating tip
⑤ Color marking band (yellow, red, blue, green)
⑥ Trimmer tool
1.800.544.2843
Probe PP006A
: 500 MHz
: 10×±2%
: 10 M (Ω) // 12 pF
: 13 pF to 23 pF
: 1.2 meters, Connector type BNC
: Yes
6
5
4
3
1
2
Item
www.calcert.com
15
10
20
5
25
0
30
PK116 accessory
PP006A
package
1
1
1
1
1
1
1
1
4×2
4×2
1
1
sales@calcert.com

Advertisement

loading

Summary of Contents for Teledyne Lecroy PP006A

  • Page 1 Probe PP006A The PP006A is a high impedance passive probe with a wide range of probing accessories. It is designed to give high impedance probing capability to a circuit when attached to a LeCroy Oscilloscope. System BW (-3dB) : 500 MHz System Attenuation : 10×±2%...
  • Page 2 Probe Adjustment LF Adjustment LF compensation is made by connecting the probe to the cal signal and adjusting the compensation trimmer in the BNC-box at end of the probe. Connect the BNC end of the probe to a channel of the oscilloscope. Connect the sprung hook end (1) and the ground lead (2) to the CAL out and GND on the scope Calibrator area.
  • Page 3 Installation for Safe and Efficient Operation Operating Environment Before using the Probe, ensure that its operating environment will be maintained within these parameters : Operation In-door Use Temperature 5 to 40 C or 41 to 104 Humidity < 80 % RH (non-condensing) The Probe has been designed to comply with EN61010-1 Installation (Overvoltage) Category II, 600 V, Pollution Degree 2.
  • Page 4 Cleaning and Maintenance The exterior of the probe and cable should only be cleaned using a soft cloth moistened with water or isopropyl alcohol. The use of abrasive agents, strong detergents, or other solvents may damage the probe. Assure that the input receptacles are free of debris before inserting connection accessories.