Teledyne Lecroy CP030A Operator's Manual

Teledyne Lecroy CP030A Operator's Manual

Current probes

Advertisement

Quick Links

Operator's Manual
CP030/CP030A
CP031/CP031A
Current Probes

Advertisement

Table of Contents
loading

Summary of Contents for Teledyne Lecroy CP030A

  • Page 1 Operator’s Manual CP030/CP030A CP031/CP031A Current Probes...
  • Page 3 CP03x/CP03xA Current Probes Operator’s Manual April 2015...
  • Page 4 Teledyne LeCroy documentation for internal educational purposes. Teledyne LeCroy is a registered trademark of Teledyne LeCroy, Inc. Windows is a registered trademark of Microsoft Corporation. Other product or brand names are trademarks or requested trademarks of their respective holders. Information in this publication supersedes all earlier versions.
  • Page 5: Table Of Contents

    Environmental Characteristics ......................5 Physical Characteristics ........................5 Safety Ratings ............................ 5 CP030/CP030A Maximum Input Current vs. Frequency ..............6 CP030/CP030A Insertion Impedance vs. Frequency ..............6 CP031/CP031A Maximum Input Current vs. Frequency ..............7 CP031/CP031A Insertion Impedance vs. Frequency ..............7 Operation ..............................
  • Page 6 Cleaning ............................16 Calibration Interval ..........................16 Service Strategy ..........................16 Troubleshooting ..........................16 Returning a Product for Service ......................17 Contact Teledyne LeCroy ........................18 Warranty ...............................18 Certifications ............................19 Declaration of Conformity — EMC ....................19 Declaration of Conformity – Low Voltage Directive ..............20 Environmental Compliance ......................20...
  • Page 7: Safety Instructions

    Operator’s Manual Safety Instructions Follow these instructions to keep this oscilloscope accessory operating in a correct and safe condition. Observe generally accepted safety procedures in addition to the precautions specified in this section. The overall safety of any system incorporating this accessory is the responsibility of the assembler of the system.
  • Page 8 CP03x/CP03xA Current Probes test before connecting or disconnecting the probe. Avoid damaging cables through excessive bending. Never install or remove the probe on bare conductors which are energized. The transformer core and shield are grounded but not insulated and may contact the conductor when the locking lever is open.
  • Page 9: Introduction

    The probes utilize a combination of Hall-effect and transformer technology which enables measurements to be made on DC, AC and impulse currents. The CP030 and CP030A have a 50 MHz bandwidth and are designed to measure continuous currents up to 30 Amp.
  • Page 10: Nominal Characteristics

    CP03x/CP03xA Current Probes Nominal Characteristics CP030 CP030A CP031 CP031A Intended Output Load 1 MΩ 1 MΩ Maximum Permitted 300 V CAT I (insulated conductor) 300 V CAT I (insulated conductor) Circuit Voltage Sensitivity * 10 mA/div to 1 mA/div to...
  • Page 11: Environmental Characteristics

    Operator’s Manual Environmental Characteristics CP030 0 ⁰C to 40 ⁰C CP030A CP031 0 ⁰C to 40 ⁰C CP031A Operating Temperature Operating RH ≤ 80% (non-condensing) -10 ⁰C to 50 ⁰C ≤ 80% (non-condensing) -10 ⁰C to 50 ⁰C Storage Temperature Storage RH ≤...
  • Page 12: Cp030/Cp030A Maximum Input Current Vs. Frequency

    CP03x/CP03xA Current Probes CP030/CP030A Maximum Input Current vs. Frequency CP030/CP030A Insertion Impedance vs. Frequency...
  • Page 13: Cp031/Cp031A Maximum Input Current Vs. Frequency

    Operator’s Manual CP031/CP031A Maximum Input Current vs. Frequency CP031/CP031A Insertion Impedance vs. Frequency...
  • Page 14: Operation

    Connecting to the Test Instrument The probe has been designed for use with the Teledyne LeCroy instruments equipped with the ProBus interface. When you attach the probe output connector to the instrument’s input connector, the instrument will: Recognize the probe model •...
  • Page 15: Operating With An Oscilloscope

    Operating with an Oscilloscope When the probe is connected to a Teledyne LeCroy oscilloscope, the displayed scale factor and measurement values will be adjusted to account for the effective gain of the probe.
  • Page 16 The front panel Volts/div knob controls the oscilloscope’s scale factor and the probe’s internal attenuation to give full available dynamic range from 10mA/div to 15 A/div for CP030 and CP031, or 1 mA/div to 15 A/div for CP030A and CP031A. CP031A and CP031A will automatically adjust attenuation based on the A/div setting.
  • Page 17: Performance Verification

    Performance Verification This procedure can be used to verify the warranted characteristics of the probe. The recommended calibration interval for the models CP030, CP030A, CP031, and CP031A Current Probes is one year. The complete performance verification procedure should be performed as the first step of annual calibration. Performance verification can be completed without removing the probe covers or exposing the user to hazardous voltages.
  • Page 18: Preliminary Procedure

    CP03x/CP03xA Current Probes Description Minimum Requirements Example Equipment Wide Band Oscilloscope 200 MHz bandwidth Teledyne LeCroy WaveRunner 6 Zi ProBus interface equipped Digital Multimeter (2) DC: 0.1% Accuracy Agilent Technologies 34401A 5½ digit resolution Fluke 8842A-09 Function Generator 50 Hz sine wave output 3 Vrms Teledyne LeCroy WaveStation 3082 into 50Ω...
  • Page 19: Functional Check

    Operator’s Manual Functional Check The functional check will verify the basic operation of the probe functions. It is recommended to perform the functional check prior to the performance verification procedure. 1. Open the C1 setup dialog and confirm that the bandwidth is set to Full (BWL OFF).
  • Page 20: Performance Verification Test Record

    CP03x/CP03xA Current Probes 1. Set the Function generator to 50 Hz sine wave, output voltage at 3 Vrms with 50Ω output. 2. Remove the probe from the instrument and reconnect using the ProBus extension cable. Connect the BNC male connector of the ProBus extension to DMM #1 using a BNC Female to Dual Banana adapter.
  • Page 21: Cp03X Test Record

    Operator’s Manual CP03x Test Record Model: _________________________________ Serial Number: _________________________________ Asset/Tracking Number: _________________________________ Date: _________________________________ Technician: _________________________________ Equipment Model Serial Number Calibration Due Date Digital Multimeter #1 Digital Multimeter #2 Function Generator * The function generator used in this performance verification procedure is used for making relative measurements.
  • Page 22: Care And Maintenance

    Service Strategy Defective probes must be returned to a Teledyne LeCroy service facility for diagnosis and exchange. A defective probe under warranty will be replaced with a factory refurbished probe. A probe that is not under warranty can be exchanged for a factory refurbished probe.
  • Page 23: Returning A Product For Service

    Operator’s Manual Returning a Product for Service Contact your local Teledyne LeCroy service center for calibration or other service. If the product cannot be serviced on location, the service center will give you a Return Material Authorization (RMA) code and instruct you where to ship the product. All products returned to the factory must have an RMA.
  • Page 24: Contact Teledyne Lecroy

    Spare parts, replacement parts and repairs are warranted for 90 days. In exercising its warranty, Teledyne LeCroy, at its option, will either repair or replace any assembly returned within its warranty period to the Customer Service Department or an authorized service center.
  • Page 25: Certifications

    2 Emissions which exceed the levels required by this standard may occur when the probe is connected to a test object. 3 To ensure compliance with the applicable EMC standards, use high quality shielded interface cables. UROPEAN ONTACT Teledyne LeCroy Europe GmbH Im Breitspiel 11c D-69126 Heidelberg Germany Tel: (49) 6221 82700 Australia &...
  • Page 26: Declaration Of Conformity - Low Voltage Directive

    Many countries prohibit the disposal of waste electronic equipment in standard waste receptacles. For more information about proper disposal and recycling of your Teledyne LeCroy product, visit teledynelecroy.com/recycle. Restriction of Hazardous Substances (RoHS) The product and its accessories conform to the 2011/65/EU RoHS2 Directive, as they...