Summary of Contents for Teledyne Lecroy PP066

  • Page 1 Operator’s Manual PP022 Passive Probe...
  • Page 2 However, clients are encouraged to duplicate and distribute Teledyne LeCroy documentation for internal educational purposes. Teledyne LeCroy is a registered trademark of Teledyne LeCroy, Inc. Other product or brand names are trademarks or requested trademarks of their respective holders. Information in this publication supersedes all earlier versions.
  • Page 3: Double Insulation

    DOUBLE INSULATION PROTECTIVE (EARTH) TERMINAL Introduction The Teledyne LeCroy PP022 is a passive high impedance voltage oscilloscope probe designed and calibrated for use with general purpose oscilloscopes having an input impedance of 1 MΩ and can be compensated for use with instruments having an input capacitance between 10 to 20 pF.
  • Page 4 PP022 Passive Probe Specifications Input Rise Compensation Imped. Time Length Range Model Attn. R (MΩ) C (pF) (MHz) (ps) (pF) PP022 10 - 20 Attenuation Ratio ..... 10:1 ± 1% Max. Input Voltage....Measurement Category I : 500 Vrms 1500 V transient overvoltage (see voltage derating curve on p.
  • Page 5 Operator’s Manual PP022 Input Impedance Profile Voltage vs. Frequency Derating Curve...
  • Page 6: General Safety Information

    PP022 Passive Probe EC Declaration of Conformity The probe is marked with this symbol to show compliance was demonstrated to the specifications listed in the Official Journal of the European Communities: Low Voltage Directive 2006/95/EC: EN 61010-031/A1:2008 Safety requirements for electrical equipment for measurement, control and laboratory use –...
  • Page 7 Operator’s Manual Do not overload. Do not apply any potential to the probe leads that exceeds the maximum rating of the probe. Always comply with the Voltage vs. Frequency Derating Curve. Observe all terminal ratings. Observe all markings on the oscilloscope before connecting.
  • Page 8 PP022 Passive Probe Probe Compensation Proper compensation of the probe is required to assure amplitude accuracy of the waveform being measured by matching the probe to the oscilloscope’s input capacitance. Compensation should be adjusted whenever the probe is connected to or transferred between oscilloscopes. Low Frequency Apply a 1 kHz square wave to the probe or connect to the oscilloscope’s calibrator output.
  • Page 9: Probe Accessories

    Operator’s Manual Over-compensated, correct, and under-compensated waveforms Probe Accessories Probe Tips Both the rigid probe tip and the spring tip can be screwed on to the probe body by hand. Be sure not to overtighten the tips to avoid damaging the threads.
  • Page 10 PP022 Passive Probe Replaceable Parts Item Part Number Description PK5-2.5MM-101 Ground Lead with Alligator Clip PK5-2.5MM-102 Rigid Probe Tip (1 installed standard) PK5-2.5MM-103 Short Collar (installed standard) PK5-2.5MM-104 Sprung Hook PK5-PK6-001 Trimmer Tool PK5-2.5MM-105 Long Collar PK5-2.5MM-106 IC Tip Insulators (Black, Brown, Blue, Green, Gray) PK5-2.5MM-107 Tip Insulator 2.5mm PK5-2.5MM-108...
  • Page 11: Warranty

    Spare parts, replacement parts and repairs are warranted for 90 days. In exercising its warranty, Teledyne LeCroy, at its option, will either repair or replace any assembly returned within the warranty period to the Customer Service Department of an authorized service center.
  • Page 12 925924-00 May, 2015...