Operation; Grounding The Probe; Handling The Probe - Teledyne Lecroy PPE6KV-A User Manual

High voltage passive probe
Table of Contents

Advertisement

Operation

The PPE6KV-A is rated for 1000 Vrms or 1500 VDC CAT II.
CAUTION: The maximum input voltage decreases as frequency increases.
Observe the Voltage vs. Frequency derating curve on p.3 and the Voltage
Rating for Accessories table on p.8.
To meet all specifications, warm the probe and instrument at least 20 minutes before
use, and always operate within the safe environmental conditions.

Grounding the Probe

Connect the probe to the oscilloscope input, then connect the ground lead to earth
ground before performing any measurements. Note that all accessible metal parts are
connected to the BNC instrument connector (GND), except for the probe tip and the BNC
center conductors.
CAUTION: The PPE6KV-A probe is designed for ground-referenced
measurements only.

Handling the Probe

To avoid any injury, handle the probe with care, especially the sharp contact tip.
Note that the probe cable is a sensitive part of the probe. Do not damage the cable
through excessive bending or pulling.
Avoid mechanical shock to this product in general to guarantee accurate performance
and protection.