7. Troubleshooting
7.1 Error codes
The transmitter's software provides a variety of self diagnostics information, for example,
FLASH and program memory checksum, probe communication status, probe checksum,
operation voltage check, and oscillator fault check.
At startup, the transmitter's software checks the factory/user settings checksum, program
memory checksum, and oscillator fault status. Other checks are made during runtime. Possible
error codes are listed in
Table 52 Error codes, descriptions, and texts
Code
1
Probe T measurement error
2
Probe RH measurement error
3
Probe communication error
4
Probe checksum error
5
Probe message form error
6
Program's flash checksum error
7
Current settings checksum error
(RAM)
8
Factory FLASH not initialized
9
User FLASH not initialized
10
Voltage is too low to operate correctly Voltage too low
11
Measurements not available
12
Oscillator fault bit active
13
Analog output quantity invalid
14
Display quantity invalid
15
Invalid relay setting
1)
Measurement parameters are referred to as 'quantities' in the transmitter's software.
You can view the error text with the errs serial line command. If you are using a transmitter
with the optional display, the error code is shown on the display in the following format:
ERR:Code-1[.Code-2][.Code-n]. Multiple errors are separated by a period (".") between the

error codes.

Table 52 (page
67).
Description
Chapter 7 – Troubleshooting
Error text
Probe T meas
Probe RA meas
Probe communication
Probe checksum
Probe message form
Program code checksum
Settings checksum
Factory defaults empty
User defaults empty
Measurements not available
HW fault 1
Analog output quantity invalid
1)
Display quantity invalid
Relay quantity invalid