CAUTIONS AND WARNINGS

To prevent injury and damage to your scale, please follow these instructions very carefully:
Do not transport the scale while platform is loaded. Be sure the scale is tilted enough to assure it is rolling properly on its
wheels when it is moved for cleaning or transported from one area to another.
For accurate weighing the scale must be placed on a flat, stable surface.
For accurate weighing, verify proper operation according to the procedure described in this manual before each use.
Do not use in the presence of flammable materials.
Operating at other voltages and frequencies than specified could damage the equipment.
If the "LOW BAT" indicator activates, for accurate weighing replace the batteries or connect the scale to an AC power source
as soon as possible.
Only authorized, qualified personnel shall service the scale.
This equipment has been tested and found to comply with the EMC limits for the Medical Device Directive 93/42/EEC (EN
55011 Class A and EN 60601-1-2). These limits are designed to provide reasonable protection against harmful interference in
a typical medical installation. The equipment generates, uses and may radiate radio frequency energy and, if not installed and
used in accordance with these instructions, may cause harmful interference to other devices in the vicinity. However, there is
no guarantee that interference will not occur in a particular installation. If this equipment does cause harmful interference with
other devices, which can be determined by turning the equipment off and on, the user is encouraged to try to eliminate the
interference by one or more of the following measures:
Reorient or relocate the receiving device.
Increase the separation between the equipment.
Connect the equipment into an outlet on a circuit different from that to which the other device(s) is connected.
Consult Customer Service or a field service technician for help.
DEFINITION OF SYMBOLS
Attention, Consult Accompanying Documents
Type BF Applied Parts
AC to DC power supply
500KLWA SCALE SPECIFICATIONS
Capacity and Resolution:
Power Requirements:
Environmental:
Platform Dimensions:
STORAGE SPECIFICATIONS
Storage Temperatures:
0°C to 40°C / 32°F to 104°F
Humidity:
10% - 95% RH
Atmospheric Pressure:
500hPa – 1060hPa
500KLWA SCALE CONNECTIVITY INFORMATION:
This scale is designed to connect to other devices via a USB port, to allow acquisition of weight, height and BMI from
the scale, to help close a source of potential medical errors caused by incorrectly copying and then recording a weight
taken. This scale is EMR-EHR compatible. For more information regarding how to connect this scale to your
computer, a monitor or other device please contact Customer Service at one of the numbers listed on the next page.
40C
Temperature range
(10°C to 40°C / 50°F to 104°F)
10C
Humidity range
(10% - 95% RH)
Atmospheric pressure range
(500hPa – 1060hPa)
220 kg x 0.1 kg / 500 lb x 0.2lb
6-AA cell 1.5V batteries or 120-240V AC – 9V DC 50-60Hz Adapter Model ADPT30
BATTERIES AND ADAPTER ADPT30 INCLUDED.
Operating Temperatures: 10°C to 40° C / 50°F to 104°F
Atmospheric Pressure: 500hPa to 1060hPa
Humidity: 10% to 95% RH
32.4 cm x 39.4 cm x 5.7 cm / 12-¾" x 15-½ " x 2 ¼"
*WARNING:
For use with Adapter Model UE 15WCP-090050SPA/ADPT30 (IEC only).
Complies with UL Std No. 60601-1. Type BF applied parts. Internally
powered.
Equipment is not suitable for use in the presence of flammable
anesthetic mixture with air or with oxygen or nitrous oxide.
3
AC to DC power supply
(9V DC, 500mA)