User Manual
PP024
Passive Probe

Advertisement

Table of Contents
loading

Summary of Contents for Teledyne Lecroy PP024

  • Page 1 User Manual PP024 Passive Probe...
  • Page 2 Administration Regulations. Export, reexport or diversion contrary to U.S. law is prohibited. Teledyne LeCroy is a registered trademark of Teledyne LeCroy, Inc. Other product or brand names are trademarks or requested trademarks of their respective holders. Information in this publication supersedes all earlier versions.
  • Page 3: Terms And Symbols

    DOUBLE INSULATION PROTECTIVE (EARTH) TERMINAL Introduction Teledyne LeCroy PP024 is a passive high impedance voltage oscilloscope probe designed and calibrated for use with general purpose oscilloscopes having an input impedance of 1 MΩ and can be compensated for use with instruments having an input capacitance between 10 and 20 pF.
  • Page 4: Specifications

    PP024 Passive Probe Specifications Input Rise Compensation Imped. Time Length Range Model Attn. R (MΩ) C (pF) (MHz) (ps) (pF) PP024 10 - 20 Attenuation Ratio ..10:1 ± 1% Max. Input Voltage ..Circuits not directly connected to Mains: 500 Vrms, .........
  • Page 5 User Manual PP024 Input Impedance Profile Voltage vs. Frequency Derating Curve EAR99 technology subject to restrictions on copyrights page.
  • Page 6: Ec Declaration Of Conformity

    PP024 Passive Probe EC Declaration of Conformity The product conforms to the applicable European Union requirements per Low Voltage Directive (LVD) 2014/35/EU. Compliance was demonstrated to the following specification as listed in the Official Journal of the European Communities: IEC/EN 61010-031:2015 Safety requirements for electrical equipment for measurement, control and laboratory use –...
  • Page 7: General Safety Information

    User Manual General Safety Information Observe generally accepted safety procedures in addition to those listed here to avoid personal injury or damage to equipment. The overall safety of any system incorporating this accessory is the responsibility of the assembler of the system. Connect only to grounded instruments.
  • Page 8 • cause deterioration of the probe body, cables, and accessories. Service Refer all repairs to qualified service personnel. Contact Teledyne LeCroy at the following address to arrange to return the probe for service: Teledyne LeCroy, Inc. 700 Chestnut Ridge Road...
  • Page 9: Probe Compensation

    High Frequency Apply a 1 MHz square wave to the probe (< 0.7 ns rise-time). Remove the plastic caps from the two, HF trimmers on the PP024 BNC Compensation Box. Adjust HF2 then HF1 until you achieve a flat-topped square wave.
  • Page 10 PP024 Passive Probe Over-compensated, correct, and under-compensated waveforms Probe Accessories Probe Tips Both the rigid probe tip and the spring tip can be screwed on to the probe body by hand. Be sure not to overtighten the tips to avoid damaging the threads.
  • Page 11: Replaceable Parts

    User Manual Replaceable Parts Item Part Number Description Quantity PK6-5MM-101 Rigid Probe Tip (1 installed standard) PK5-PK6-001 Trimmer Tool PK6-5MM-102 Sprung Hook PK5-PK6-002 Color Coding Rings (2 ea. yellow, red, green, blue) PK6-5MM-103 Ground Lead with Alligator Clip PK6-5MM-104 Ground Spring PK6-5MM-105 PCB Adapter PK6-5MM-106...
  • Page 12: Warranty

    Spare parts, replacement parts and repairs are warranted for 90 days. In exercising its warranty, Teledyne LeCroy, at its option, will either repair or replace any assembly returned within the warranty period to the Customer Service Department of an authorized service center.