MAINTENANCE

Cleaning

Clean the outside of the MDA800A using a soft cloth moistened with water
or isopropyl alcohol solution. Do not use harsh or abrasive cleansers.
Dry thoroughly before using. Do not submerge the instrument or allow
moisture to penetrate it.
CAUTION. Do not attempt to clean internal parts.

Language Selection

To change the language that appears on the instrument touch screen, from
the menu bar, choose Utilities > Preference Setup > Preferences and make
your Language selection.
You can also select at power on by touching the Language icon
when it appears at the upper-right of the touch screen.
Reboot the MDA800A after changing language.

Adding Option Keys

Software options are availble to extend the capabilities of an MDA800A. To
purchase an option, contact your Teledyne LeCroy sales representative. You
will receive a license key via email that activates the optional features on the
instrument. To install the key:
1. From the menu bar, choose Utilities > Utilities Setup > Options.
2. Touch Add Key.
3. Enter the new license key and click OK.
4. Reboot the instrument.
See the product page on our website for a full list of options.
44
929958-00 Rev A

Calibration

The MDA800A is calibrated at the factory prior to being shipped. This
calibration is run at 23° C (± 2° C) and is valid for temperatures ± 5° C of
the original calibration temperature. Within this temperature range, the
MDA800A will meet all specifications.
Warm up the MDA800A for at least 20 minutes prior to use. During the
warm-up period, the instrument will automatically initiate calibrations to
ensure that it is always calibrated.
When the MDA800A is used outside of the factory calibration
temperature range, a user-invoked temperature dependent calibration is
recommended. There are two options for this calibration: Calibrate All or
Calibrate Current Setting.
Calibrate All - All possible combinations of vertical and horizontal
settings are calibrated at the current temperature. This calibration is valid
for the current temperature ± 5° C and takes about 50 minutes.
Calibrate Current Setting - The instrument is calibrated at the current
vertical and horizontal settings only. This calibration is valid for these
settings at the current temperature ± 5° C and takes under 30 seconds.
It is also recommended that the MDA800A be calibrated when it has
been more than one month since the previous calibration.
CAUTION. Remove all inputs from the instrument prior to
calibration.