SET UP

Safety

To maintain the instrument in a correct and safe condition, observe gener-
ally accepted safety practices in addition to the precautions specified here.
Symbols
CAUTION of potential damage
to equipment, or WARNING of
potential bodily injury. Refer to
manual. Do not proceed until the
information is fully understood
and conditions are met.
WARNING. Risk of electric shock
or burn.
Electrostatic Discharge Warning.
Operating Environment
Temperature
5 °C to 40 °C
Humidity
Maximum RH 80% (non-condensing) up to 30 °C
decreasing linearly to RH 50% at 40 °C
Altitude
Up to 10,000 ft (3,048 m) at or below 30 °C
Use indoors only.
Use only within the operational environment listed. Do not use in wet or
explosive environments.
4
933197-00 Rev A
Frame or chassis terminal
Alternating current
Power On/Standby Power
Power and Ground
AC Power
100-240 VAC (±10%) at 50/60/400 Hz (±5%)
Automatic AC Voltage Selection
Consumption
Maximum Operating
40 W / 40 VA
Standby
2 W / 2 VA
Use only the Power button to power down. Do not power down by
switching off a connected battery/power strip or pulling the AC line cord
from the outlet while the instrument is turned on.
CAUTION. In "Standby Power" mode the instrument is still
connected to the AC supply. The only way to fully power down the
instrument is to physically disconnect the power cord from the AC
outlet. We recommend unplugging the instrument if it will remain
unused for a long period of time
Use only the power cord shipped with the instrument and certified for
the country of use.
The instrument is provided with a 10A/250V 18AWG rated grounded cord
set containing a molded three-terminal polarized plug and a specific
IEC-60320 (Type C13) connector for making line voltage and safety
ground connections.
Maintain ground. The AC inlet ground is connected to the frame of the
instrument. For adequate protection against shock, connect line cords
only to outlets with safety ground contacts.
WARNING. Interrupting the protective conductor inside or outside
the WavePulser, or disconnecting the safety ground terminal,
creates a hazardous situation. Intentional interruption is prohibited.