ZD200 Differential Probe
Care and Maintenance
Cleaning
The exterior of the probe and cable should be cleaned, using a soft cloth
moistened with water. The use of abrasive agents, strong detergents, or other
solvents may damage the probe. Always ensure that the input leads are free
of debris.
CAUTION: The probe case is not sealed and should never be
immersed in any fluid.
Calibration Interval
The recommended calibration interval is one year. The Performance
Verification procedure should be performed as the first part of calibration.
Service Strategy
The ZD200 probe utilizes fine pitch surface mount devices. It is therefore
impractical to attempt to repair in the field. Defective probes must be
returned to a Teledyne LeCroy service facility for diagnosis and exchange.
Defective probes under warranty are repaired or replaced. A probe that is not
under warranty can be exchanged for a factory refurbished probe for a
modest fee. You must return the defective probe in order to receive credit for
the probe core.
Replacement Parts
The probe connection accessories and other common parts can be ordered
through the North America Customer Care Centers.
Replacement Part
Accessory Kit
Y-Lead Adapter
6
Part Number
PACC-ZD007
PACC-ZD008