Page 4
Teledyne LeCroy documentation materials is strictly prohibited. Teledyne LeCroy is a registered trademark of Teledyne LeCroy, Inc. Other product or brand names are trademarks or requested trademarks of their respective holders. Information in this publication supersedes all earlier versions.
Performance Verification ............... 7 Required Test Equipment ................. 7 Preliminary Procedure ................9 Functional Check ..................9 Verification Procedure................10 ZD200 Test Record ................. 11 Reference Material ................12 Certifications ................... 12 Returning a Product for Service ............13 Technical Support .................. 15...
Page 6
Spare parts, replacement parts and repairs are warranted for 90 days. In exercising its warranty, Teledyne LeCroy, at its option, will either repair or replace any assembly returned within its warranty period to the Customer Service Department or an authorized service center. However, this will be done only if the product is determined by Teledyne LeCroy’s examination to...
Operator’s Manual Safety Instructions This section contains instructions that must be observed to keep this oscilloscope accessory operating in a correct and safe condition. You are required to follow generally accepted safety procedures in addition to the precautions specified in this section. The overall safety of any system incorporating this accessory is the responsibility of the assembler of the system.
With the ProBus interface, the ZD200 probe becomes an integral part of the oscilloscope, able to be controlled from the oscilloscope’s front panel. The oscilloscope provides power to the probe, so there is no need for a separate power supply or batteries.
Operator’s Manual Standard Accessories The ZD200 probe is provided with numerous accessories to make probing different test points easier than ever. The probe is also supplied with an Operator’s Manual and a Certificate of Calibration. Accessory Quantity Description Hook Clips...
Using one of the available accessories makes the ZD200 probe with its small profile and low mass head ideally suited for applications in dense circuitry.
DC offset from the probe's amplifier output to improve the measurement accuracy. Auto Zero is invoked manually from the ZD200 dialog that appears when the probe is connected to the oscilloscope. Always perform Auto Zero after the probe is warmed up (recommended warm-up time is 20 minutes).
Verification procedure should be performed as the first part of calibration. Service Strategy The ZD200 probe utilizes fine pitch surface mount devices. It is therefore impractical to attempt to repair in the field. Defective probes must be returned to a Teledyne LeCroy service facility for diagnosis and exchange.
This procedure can be used to verify the warranted characteristics of the ZD200 High Impedance Active Probe. It tests LF Attenuation Accuracy. The recommended calibration interval for the ZD200 probe is one year. Complete the performance verification as the first step of annual calibration.
Page 14
ZD200 Differential Probe List of Required Test Equipment. Description Minimum Requirement Example Equipment Digital Oscilloscope ProBus Interface Teledyne LeCroy Windows with software HDO6000, version 6.4.1.5 or later WaveRunner 8000, WavePro 7 Zi-A Digital Multimeter (DMM) 4.5 digit Keysight 34401A with test probe leads DC: 0.1% Accuracy...
LeCroy representative or visit teledynelecroy.com if the software in your oscilloscope requires updating. 1. Connect the ZD200 probe to the female end of the PROBUS-CFO1. Connect the male end of the PROBUS-CFO1 to oscilloscope C1. 2. Turn the oscilloscope on and allow at least 30 minutes warm-up time for the ZD200 and test equipment before performing the Verification Procedure.
10. Install the straight tips on the ZD200 inputs. 11. Connect the signal generator output to the ZD200 inputs using the BNC to mini-grabber cable. 12. Read the voltage from the DMM and record this value on the test data sheet.
Operator’s Manual ZD200 Test Record Technician:_____________________________________________ Date:_____________________ Equipment Used Item Model Serial Number Cal Due Date Oscilloscope Function Generator Digital Multimeter Probe Lead Test Record, LF Attenuation Accuracy Step Description Results DMM AC Voltage DMM AC Voltage/10 DMM AC Voltage Gain Error, (Test Limit ≤...
ZD200 Differential Probe Reference Material Certifications Teledyne LeCroy certifies compliance to the following standards as of the date of publication. For the current certifications, see the EC Declaration of Conformity shipped with your product. EMC Compliance EC D - EMC...
Page 19
The probe is subject to disposal and recycling regulations that vary by country and region. Many countries prohibit the disposal of waste electronic equipment in standard waste receptacles. For more information about proper disposal and recycling of your Teledyne LeCroy product, visit teledynelecroy.com/recycle. ESTRICTION OF...
Extended warranty, calibration and upgrade plans are available for purchase. Contact your Teledyne LeCroy sales representative to purchase a service plan. For a complete list of Teledyne LeCroy offices by country, including our sales and distribution partners, visit: teledynelecroy.com/support/contact.
Operator’s Manual Technical Support Live Support Registered users can contact their local Teledyne LeCroy service center at the number listed on our website. You can also request Technical Support via the website at: teledynelecroy.com/support/techhelp Resources Teledyne LeCroy publishes a free Technical Library on its website. Manuals, tutorials, application notes, white papers, and videos are available to help you get the most out of your Teledyne LeCroy products.
Page 22
ZD200 Differential Probe THIS PAGE INTENTIONALLY LEFT BLANK.