CALIBRATION

This scale is factory calibrated and does not require calibration prior to use.
The calibration procedure may be performed using a known/certified weight in either pounds or
kilograms. The default is in pounds. The following instructions assume the scale is set to pounds. If
you prefer to calibrate in kilograms, please call Customer Service at 1-800-815-6615.
Note: Calibration weight must be between 60 lb (27 kg) and 300 lb (150 kg).
Turn the scale on. While "START" is displayed (for
approximately 3 seconds), press ZERO - UNITS - UNITS - ZERO
Press SEND
Press SEND to calibrate with 300 lb. If calibrating with less
weight, press WEIGHT/HEIGHT/BMI to change the weight and
then use the arrows to change the flashing digit; press
WEIGHT/HEIGHT/BMI to move to the next digit, until you reach
the desired weight (between 60 lb and 300 lb)
Clear the platform and press SEND
Wait without touching or shaking the scale until the zero
calibration process is finished.
Load the platform with the required weight and press SEND
Wait without touching or shaking the scale until calibration
process is finished.
Remove weight from the platform.
CALIBRATION ERROR LIST
ERROR
NO.
Err 1
Mathematical mistake; division by zero.
Cannot calculate calibration factor
Err 2
Weight loaded too low. Cannot
calibrate
Err 3
Unstable weight. Cannot calibrate
Err 5
Calibration disabled
Err 6
High saturation state (high A/D)
Err 7
Low saturation state (low A/D)
ACTION
DEFINITION
"CAL" will appear on the LCD
Scale will reboot. "StArt" will
appear followed by the software
CHECK/REPLACE
Will show when trying to calibrate with no
calibration weight
While in calibration procedure add more weight in
load mode
Check the load cells' surroundings and ensure
that nothing touches them and the cables are
well connected
On the back of the display, locate the silver
extended screw and plastic tab. Use a nut driver
or pliers to remove the tab and gain access to the
push button underneath. Press and release one
time to enable proper calibration. When finished,
toggle the calibration switch to enable.
The load cell cable is not connected correctly.
Check the cables and mechanical connection.
See Err 6
24
DISPLAY SHOWS
"Load/300.0 lb"
"CLEAR"
Zero Calibration
"-----------"
Weight calibration
"Put XXX.X"
"-----------"
"DONE"/"CLEAR"
version and "0.0"